![JD HEALTHCARE LEVABO TURN ALL User Manual Download Page 16](http://html1.mh-extra.com/html/jd-healthcare/levabo-turn-all/levabo-turn-all_user-manual_3556104016.webp)
16
Turn All userguide 3C EN 2021.05.21 Rev.4
Any product malfunction that has
caused personal injury must be notified
immediately to the appropriate statutory
authority and Levabo ApS.
Note!
The competent authority may require
an assessment by an independent expert,
including a written report. The expert
shall be selected in agreement with the
competent authority.
The technical safety assessment includes
the following elements:
•
What the event can be attributed to
•
Whether the product was in good
condition
•
If there is any further danger after repair
of malfunction
•
Whether new knowledge has been
obtained, that requires further
investigation or the introduction of
additional precautions.
Dimensions (L x W x H)
Cover ................................................200 x 85 x 15cm
Lifting mattress .................................... 200 x 85cm
Pump ................................................ 8 x 27 x 13.5cm
Weight
Pump ..................................................................... 1.5kg
Turn All Cover (Upper & Base) with Multi-
Patient Inner Lifting Cells .................................. 4kg
Lifting angle
Dynamic (automated) ..........................................30º
Static (manual) .............................................. 0º - 30º
Electrical data
Mains voltage ......................................................240V
Frequency ..............................................
50Hz / 60Hz
Power consumption ................................Max. 12W
Classification
.................................................. Class II
Fuse ...................................................T500mA L250V
Storage
-10º to 50º Celcius.
Relative humidity 0% to 95% without
condensation.
Conditions of use
Designed for continuous operation.
Operating temperature:
0° to 40° Celcius.
Relative humidity 20% to 95% without
condensation.
Standards
See approvals on page 16.
NOTIFICATION
OF ACCIDENTAL EVENTS
TECHNICAL DATA