Cleaning the Turn All
®
system
Turn All
®
Cover
Detach the blue PU coated
two way stretch fabric Cover Upper from
the 600D Black Dacron Cover Base, then
remove the Multi Patient Inner Lifting Cells.
The Cover Upper can be machine washed
at 90° C and tumble dried on the lowest
setting. See page 15.
Turn
All
®
Cover Upper can be cleaned in
autoclave v. Max. 121° C.
The Turn
All
®
system must be disconnected
from the power outlet before cleaning.
If the Turn All
®
system has been
contaminated by body fluids, the following
method of cleaning is recommended.
In addition to machine washing the Turn All
Cover Upper, as per previous instructions
on this page, the Turn All Multi Patient Inner
Lifting Cells can be hand cleaned in warm
soapy water. A disinfectant cleaning agent,
with a 1.0% chlorine content, may also be
used to clean both the lifting cells and the
Cover Base.
The external surfaces of the Turn All
®
Pump 3C can be sprayed with a suitable
disinfectant or wiped over with alcohol /
propyl alcohol on a damp cloth.
Note!
Do not use caustic or abrasive
cleaners or sharp objects to clean the Turn
All
®
system.
CLEANING
AND DISINFECTING