LEVABO

® 

TURN ALL

®

User Manual

Patent Pending

Summary of Contents for LEVABO TURN ALL

Page 1: ...LEVABO TURN ALL User Manual Patent Pending...

Page 2: ...instructions before attempting to use this device These operating instructions are part of the product and must be kept close to the system so that it can be consulted at any time by personnel and ope...

Page 3: ...t s mattress 6 Important information Regarding the hose connection 7 Starting and adjusting the Turn All system 8 Using the Turn All system for static side lying 9 Indications and contraindications 10...

Page 4: ...ecommended will vary according to the needs of each individual but when this form of pressure care is indicated the frequency is normally a 1 5 to 2 0 hour interval for each change of direction The au...

Page 5: ...one side is inflated the patient is gently turned 30 towards the opposite side of the mattress and then lowered again as the chambers gradually deflate After a short delay the process repeats on the o...

Page 6: ...system so that the hose connection B is at the foot end of the bed Figure 1 2 Attach the Turn All system to the bed base using the straps H on the cover Figure 1 2 Put the mattress back onto the bed o...

Page 7: ...connection B is marked with the letters R and L to indicate which side of the mattress will begin to lift first To commence turning the patient on the right side of the bed seen from the foot end the...

Page 8: ...the patient will be gently rolled onto their side up to 30 and back down again over a 30 60 or 90 minute interval followed by 15 minutes in the supine position The process is then repeated on the oppo...

Page 9: ...n the pump and continue running it until the patient has been turned to the desired side lying angle Then unplug hose connector B Figure 5 and seal it with cover E Figure 6 Finally turn off the pump C...

Page 10: ...mendations of the medical practitioner responsible for the patient s care The system is designed for patients with a body weight of up to 317 kg The Turn All system is suitable and recommended For peo...

Page 11: ...Turn All system is safe to be used by the same patient for an extended period of time The Turn All system is an aid for the prevention of pressure injuries as well as being an auxiliary therapy for p...

Page 12: ...disconnected from the power outlet during operation or there is a power failure an audible alarm will sound for approx 5 minutes The audible alarm can be silenced by pressing the Power On Off button D...

Page 13: ...month guarantee from the invoice date During this period faulty parts will be repaired or replaced free of charge if caused during the manufacturing process Levabo ApS assumes no liability for damage...

Page 14: ...he Turn All system has been contaminated by body fluids the following method of cleaning is recommended In addition to machine washing the Turn All Cover Upper as per previous instructions on this pag...

Page 15: ...ient Inner Lifting Cells H See Figure 1 Pull the connection tubes B for the Multi Patient Inner Lifting Cells through access sleeve in the Cover Upper Figures 1 2 to completely separate it from the Co...

Page 16: ...her new knowledge has been obtained that requires further investigation or the introduction of additional precautions Dimensions L x W x H Cover 200 x 85 x 15cm Lifting mattress 200 x 85cm Pump 8 x 27...

Page 17: ...A1 2011 Low Hazard Method of Test BS EN 597 1 1995 BS EN 597 2 1995 PASS EN 1021 1 2 PASS Cover bottom part PVC Black 600D PVC coating dacron fabric Pump Tested and approved in accordance to IEC 60601...

Page 18: ...ee cycle Pump 3C 317kg REPLACEMENT PARTS Product Name Product Code Size Description SWL Turn All Lifting Cells for Multi Patient System PCLV 70113 200cm L x 85cm W x 15cm H Nylon inflatable lifting ce...

Page 19: ...of manufacture Product number NOTES Declaration of Conformity CE The company Levabo ApS Sverigesvej 20A 8660 Skanderborg declares that the products are designed and manufactured in accordance with th...

Page 20: ...PC JDHG090322 Unit 3 22 Beaumont Rd Mt Kuring Gai 2080 NSW Australia 1300 791 404 02 9482 2133 www jdhealthcaregroup com au sales jdhealthcare com au Pressure Care TURN ALL QUICK START VIDEO...