15
AIR PRESSURE STATUS
This indicator shows when the air pressure is normal or too low. When the air pressure
is correct, the
“Normal”
lamp will be lit. If the
“Low”
lamp lights up, the air pressure is
too low and the mattress should not be used.
STATIC/DYNAMIC
Press this button to select
“Static”
mode or
“Dynamic”
(alternating air pressure)
mode. When the adjacent
“Static”
lamp is lit, the mattress is in static mode. When the
adjacent
“Dynamic”
lamp is lit, the mattress is in alternating pressure mode and the
cycle time is 10 minutes.
SPECIFICATIONS
Pump OC-20
Specifications
Size
28.0 x 11.0 x 20.5cm
Weight
2.6 kg
Pressure range
20 ~ 70mm Hg
Cycle time
10 minutes
System
Specifications
Classification
Class II; IPXO, No AP/APG; Type BF
Applied part
Mattress
Power rating
~230V AC/50 Hz
1A, 12W
Fuse rating
250V, T 1 A
Operating environment
Temperature
Use: 10° ~ 35° C
Storage: -15° ~ 50° C
Transport: -15° ~ 70° C
Humidity
Use: 20% ~ 80% without condensation
Storage: 10% ~ 90% without condensation
Safety standard
CE-marked