INSPECTION AND TROUBLESHOOTING
INSPECTION OF YOUR OPTICELL 2 NX AIR MATTRESS SYSTEM
An inspection plan should be established
in order for your air mattress to function and last
for as long as possible. Please contact us if you would like assistance with this. Järven Health
Care’s contact details can be found on the back cover of this guide.
INSPECTION OF THE MATTRESS AND PUMP
• Check the electrical cord and plug for signs of damage or abnormal wear.
• Check the mattress surface for signs of damage or abnormal wear.
• Check that the cells of the mattress are correctly connected to the air hoses.
• Start the pump and examine the air flow from the pump (with the air hoses
disconnected). The air flow should alternate between the two outer holes each cycle
time. The middle hole should release no air.
• Check the air hoses for signs of damage or abnormal wear.
• Check the air hose connector (see page 10).
INSPECTION OF THE OPTITEX COVER
Regularly check the cover for visible signs of damage or wear,
such as tears, cracks, holes,
stains or discolouration. Check all sides including the underside.
TROUBLESHOOTING
The pump does not work or does not start:
• Make sure the electrical plug is properly inserted, then turn on the pump’s power switch.
• If the power indicator is off, there may be a problem with the electrical socket. Test
another power outlet.
• If the power indicator is on but the pump does not work, contact Järven Healthcare.
The mattress does not inflate properly (pressure too low):
• Set the pressure to maximum, check if the air hoses are pinched anywhere or if there is
air leakage from any cell. Check that the CPR valve is closed.
Poor air flow:
• A dirty air filter can reduce air flow. Wash the air filter
with mild detergent. The air filter is
easily accessible on the back of the pump and can be released with a simple motion.