10
AIR HOSE CONNECTION
Connecting the air hoses to the pump:
To remove the protective caps and to connect or disconnect the air hose, push in the two
contact clips on the side of the connector. This action releases the hooks that hold the
connector in place. The protective caps should be put back on the connector if the air hose is
disconnected from the pump.
To minimise the risk of air leakage during hose connection, check:
• That the two hooks are not damaged and their clips are intact.
• That the black o-rings on the connector’s three air hoses are not cracked or missing.
• That no debris has entered the connector causing a bad connection or blocking the air
hoses.
The air hose connector is
fitted by a simple hand
motion. No special tools
are required.
The contact clips are on
the sides of the connector.