(II)
JWL 6
5-2 Description of Parameter Values
1.
P00:
Offset value
Displays the offset value and the keypad testing can be conducted
2.
P01:
Backlight mode
Off
:
No backlight
Auto
:
Auto on with items greater than 9d placed on the weighing pan or any key is
pressed. But auto off after N seconds (N=2s, 5s, 10s, 20s, ever) with no action
Ever=
It is always on when the weights over 9e
On
:
Backlight on
3.
P02:
Auto-off
Off
:
Non power off
5
,
10
,
30
,
60
(minutes)
4.
P03:
Unit setting
Init
: Press key
Unit
to
select the default unit when powering on the scale
:
pcs,
斤
, lboz,
g, kg, final .(final=keep the final being used unit when power off )
Use:
Press key
Unit
to select the weighing unit.
on
:
Enable the unit
off
:
Disable the
unit
5.
P04:
Zero range
d0
,
d1
,
d2
,
d3
,
d4
and
d5
. (d= scale division)
6.
P05:
Hold function
HoLd – 0
:
no hold function
HoLd – 1
:
Peak hold. Press any key to release
HoLd – 2
:
Hold after stable. Press any key to release
HoLd – 3
:
Hold after stable. Release after moving away the article
HoLd – 4
:
Press key
HOLD/LO/-
to hold. Press any key to release
7.
P06:
Check weighing memory
on
:
Check weighing on
off
:
Check weighing off
8.
P07:
Check weighing function
on:
Check weighing under the condition that the weight is within the limits and the stable
indication appears