9. Troubleshooting and Error Message
Error Message
Problem
shootings
ERR0
Exceed the zero range
The item should be within 2% of full load
ERR2
Exceed the initial zero point
1. Check whether there are other alien
articles on the scale pan, remove those
articles.
2. LOAD CELL failure, which requires to be
changed or to contact our Service.
ERR3
Exceed the A/D resolution
range
1. Check whether it is A/D failure, if yes,
please replace AD.
2. LOAD CELL failure, replacement is
required or contact our Service.
ERR4
EEPROM failure
Re-sold EEPROM or contact our Service.
ERR5
Overload condition
Remove weight that is greater than the scale
capacity from the pan.
ERR6
Exceeds the display range
--------------------------------------------
ERR7
Accumulated
number
of
weighments
exceeds
the
display range
Delete the exceeding weighments
ERR8
Lower limit is higher than upper
limit
re-act the upper limit setting and lower limit
setting
ERR9
Exceed tare or pre-tare range
The tare value should be over zero and less
than or equal to full load.
ERR10
Wrong calibration weights
Place the right weights( the calibration value
≤
full load)