4. After the Diagnostics screen opens, clamp the meter in place by pressing
Clamp
in the lower
right screen corner.
5. Press the clamp buttons on both sides of the SNAP prover.
6. After the Intelis Gas Meter is clamped to the prover, turn on the appropriate nozzles in the
center of the screen to exercise the meter.
Note:
Typically, Nozzle Number 2-245 CFH is used to exercise the Intelis Meter.
7. Exercise the meter for 10 to 15 minutes prior to proving the meter.
8. Select
Close Nozzles
.
9. Unclamp the meter.
10. Exit Special Functions.
Setting Up the Prover
1. Log on to the prover.
Note:
If you are proving an Intelis Gas Meter for the first time on the prover or if you are
changing proving parameters for the Intelis Gas Meter, you must log on as a Supervisor.
To set up Itron Intelis Gas Meter and ITRON-RTY for the first time, follow steps 2 through
10.
2. Select
Special Functions
.
08 July 2020
815-0154-00 REV 000
Itron, Inc. Page 21 of 66
Intelis Gas Meter Proving and Setup Guide
4 Proving the Intelis Gas Meter Using the Honeywell SNAP Proving
System