2. Connect the Intelis Gas Meter end of the Measurement Systems pulse prover cable to the
Intelis Gas Meter by firmly pressing the aluminum round threads of the cable into the port.
Begin by positioning the cable at the 4:30 o'clock position. Tighten the cable to the 6:30
o'clock position.
Caution:
Do not over tighten this, as doing so could damage the index housing or
the proving cable head.
Note:
The Intelis Gas Meter automatically goes into Test Mode (Menu 6) after the cable
is properly connected to the optical port.
3. Enter the Intelis Gas Meter serial number.
08 July 2020
815-0154-00 REV 000
Itron, Inc. Page 46 of 66
Intelis Gas Meter Proving and Setup Guide
5 Proving the Intelis Gas Meter Using the Measurement Systems
Proving System