Function
and Features
Copyright ©ITECH Electronic Co., Ltd.
37
9. Set loading time of step 2. If 2S is required, directly press 2 key. Press
[Enter]
key.
EDIT PROGRAM
SEQ02 On Time=2S
10. Set unloading time of step 2. If 2S is required, directly press 2 key. Press
[Enter]
key.
EDIT PROGRAM
SEQ02 Off Time=2S
11. Set test delay time of step 2. If 1S is required, directly press 1 key. Press
[Enter]
key. Tpf is delay time before measurement.
EDIT PROGRAM
SEQ02 P/F Delay Time=1S
Tpf is delay time before measurement.
12. Set conditions for stop test. COMPLETE means to stop after finished all
tests and FAILURE means to stop in case of test error. Press
[Enter]
key.
PROGRAM
Complete-Stop Failure-Stop
13. Determine whether to link to next group of test file. If it is to link to second
group, press 2 key. 0 means not to link to other test files. Press
[Enter]
key.
PROGRAM
Chain Program File=0(0-10
)
Correspondence Table of Auto test files and step parameter saving.
Program 1 Sequence
1
2
3
4
5
6
7
8
9
10
Save Group
1
2
3
4
5
6
7
8
9
10
Program 2 Sequence
1
2
3
4
5
6
7
8
9
10
Save Group
11
12
13
14
15
16
17
18
19
20
:
:
Program 10 Sequence
1
2
3
4
5
6
7
8
9
10
Save Group
91
92
93
94
95
96
97
98
99 100
14. Save the programmed files in EEPROM. A total of 10 groups of files can be
saved. If it is to save edited files in group 1, press 1 key. Press
[Enter]
key.
PROGRAM
Save Program File=1
(
1-10
)
15. Press
[ESC]