CONFigure:TEMPerature:RTD <type>,<standard>,<con>,<current mode>,<root 2>
Sets the selected channel for an RTD measurement according to the following parameters.
<
type
>
Pt25, Pt100
<
standard
>
1
Invalid (Obsolete)
2
Invalid (Obsolete)
3
(EN60751)
4-23
(User Probe 1-20)
<
con
>
3, 4
(3 or 4 wire connectivity)
<
current mode
>
+I, -I, AVE
<
root 2
>
0, 1, Off, On
Terminates the current measurement cycle and scanning routine.
Sets the scanner and data logger modes to OFF.
CONFigure?
Used to query the current configuration. The response is an ASCII string of either of the following forms:
“<
channel
>,TC
,<type>,<rj mode>,<ext rj mode>”
“<
channel
>,RTD
,<type>,<standard>,<con>,<current mode>,<root 2>”
Parameter values are as per the tables above. Does not affect the current measurement.
MEASure Group
These work similarly to the CONFigure commands, except that an immediate measurement is performed with the
result being sent to the output buffer. SINGle mode is set, and only one reading is taken. As with CONFigure, the
current measurement cycle will be aborted.
MEASure:TEMPerature:TC? <
type>,<rj mode>,<ext rj standard>
MEASure:TEMPerature:RTD? <
type>,<standard>,<con>,<current mode>,<root 2>
MEASure:CHANnel? <
channel>
Parameters are as for the CONFigure command.
SENSe Group
SENSe:TEMPerature:UNIT
<parameter>
Sets the temperature units to be used in all responses.
<parameter>
C, F, K
(degrees)
SENSe:TEMPerature:UNIT?
Queries the temperature unit in force. Responds with “C”, ” F” or “K”.
SENSe:TEMPerature:RESolution
<resolution
>
Sets the temperature resolution to be used in all responses.
<resolution>
1, 0.1, 0.01, 0.001