Isotech TTI 7 PLUS User Maintenance Manual/Handbook Download Page 17

 

Page 17 of 44 

TTI-7 Iss.02 – 06/11 

CAUTION: When measuring thermocouples with the TTI 7

PLUS 

it should be remembered that the temperature may 

be read to 0.01

C, and this is equivalent to approximately 0.4

V per digit with a type K thermocouple.  Care must 

therefore be taken to avoid exposing the instrument inputs to draughts or a heat source such as sun shining on the 

front  panel.    To  obtain  the  best  results  the  instrument  should  be  placed  in  a  temperature  controlled  room  and 

switched on for at least half an hour before measurements are made.  Running on battery power without the mains 

supply connected will also be of benefit.  When connected to the mains the batteries will be charged and additional 

heat generated inside the instrument.  Great care should also be taken to ensure that the correct materials are used 

when making connections and avoid unnecessary joints of different materials as these again will cause small errors. 

We offer a range of thermocouple connectors for use with the  TTI 7

PLUS  as

 well as a range of automatic reference 

junctions  and  an  external  reference  enclosure  with  integral  Pt100  sensor;  please  ask  our  sales  staff  for  more 

information. 

CALIBRATION 

 

The  TTI  7

PLUS 

has  been  designed  to  give  long  service  and  permanence  of  calibration,  together  with  a  low  cost  of 

ownership. Attention to detail in the design stage has produced an instrument that is quick and simple to recalibrate 

with  the  minimum  of  equipment.  Prior  to  calibration  the  unit  should  be  placed  in  a  temperature  controlled 

environment for a minimum of 4 hours.  Care could be taken to use good quality test leads, thus avoiding any thermal 

emf,  together  with  resistance  standards  of  known  value.    To  select  the  Cal  option,  press  the  MEM  key and then 

select the Cal option from the menu. The calibration counter number will be displayed.  A second menu choice will 

be given - Cal or Prt. The Prt option will print the calibration constants to the screen.  This is used as a diagnostic 

tool to ensure that there is no corruption of these constants. Select the Cal option and you now enter the calibration 

routine. The display will ask for a passcode to be entered, the default passcode number is 9252, enter this number 

and press OK.  The display will then give you the option of changing the passcode to your own personal number, if 

you  select  Yes  you  will  then  be  asked  to  enter  your  chosen  number  via  the  keypad  ending  by  pressing  OK.  Any 
number  between  0  and  9999  is  valid.  You  will  then  be  asked  to  confirm  your  number.  Once  the  new  number  is 

accepted the calibration counter will increment 1.  The display will then give a choice of calibration RTD, T/C, RJA, 

RJB, it is preferable to calibrate the RTD ranges first.  

RTD Calibration 

 

Select this option by pressing RTD.  Select Pt100, the display will then prompt you to connect 100 ohm resistance 

standard  to  channel  A.    When  connected  press  OK  and  you  will  be  prompted  to  enter  the  exact  value  of  the 

standard. Once the value is entered and the OK pressed you will be asked to confirm the value by pressing OK again.  

The instrument will then calibrate itself against this standard storing the calibration constants in the memory.  Next 

you will be asked to connect a 400 ohm resistance standard and repeat the above.  Finally you will be prompted to 

connect a 250 ohm resistance standard and again repeat the above.  Once this calibration is completed the display 
will return to the main menu.  You now have the option to calibrate the Pt25 range.  Follow the display instructions 

using the 100Ω 25.5 Ω and 62.5 Ω standards. 

Thermocouple Calibration 

 

Select this option by pressing the T/C key in the calibration menu.  Connect an mV source to the thermocouple input 

of channel A using low thermal copper leads and connectors.  The display will prompt you to set the input to 0mVdc 

and press OK.  Next +75mVdc should be supplied and OK pressed.  Finally -75mVdc should be supplied and again 

the OK pressed.  These three values will calibrate the complete thermocouple range of the instrument.  The display 

will again return to the main calibration screen and the reference junction should now be calibrated. 

Summary of Contents for TTI 7 PLUS

Page 1: ...4 0 1704 543830 Fax 44 0 1704 544799 Internet www isotech co uk E mail info isotech co uk The company is always willing to give technical advice and assistance where appropriate Equally because of the...

Page 2: ...TEM REQUIREMENTS 19 Development 19 License 19 INSTALLING CAL NOTEPAD 19 STARTING CAL NOTEPAD 20 Configuring for the TTI 7PLUS 20 REMOTE CONTROL INTERFACES 21 INTRODUCTION 21 Version History 21 RS 232...

Page 3: ...AND SUMMARY 41 Programming Language 41 Measurement Configuration Commands 41 Triggering Commands 41 Routing Commands 42 Memory Commands 42 Datalogger Commands 42 Mathematical Operation Commands 42 Aut...

Page 4: ...UARANTEE INTERFERENCE WITH OR FAILURE TO PROPERLY MAINTAIN THIS INSTRUMENT MAY INVALIDATE THIS GUARANTEE RECOMMENDATION The life of your ISOTECH Instrument will be prolonged if regular maintenance and...

Page 5: ...must be correctly earthed This equipment is a Class 1 Appliance A protective earth is used to ensure the conductive parts can not become live in the event of a failure of the insulation The protectiv...

Page 6: ...munity Signal Lines prEN60 1000 4 4 500V Radiated Immunity prEN60 1000 4 3 3V m Electrostatic Discharge Immunity prEN60 1000 4 2 8kV Electrical Safety Standard IEC1010 1 Amend 1 INSTALLATION The TTI 7...

Page 7: ...arger will automatically switch to trickle charge or fast charge mode depending upon the battery state LOW BATTERY indication is displayed when the batteries have approximately 10 charge left When thi...

Page 8: ...5 C Pt25 200 to 100 2 5 to 15 1 mA 0 001 0 02 C Pt25 100 to 500 15 to 75 1 mA 0 001 0 01 C Pt25 500 to 670 75 to 115 1 mA 0 001 0 02 C Pt100 200 to 100 10 to 60 0 5 mA 0 001 0 02 C Pt100 100 to 500 60...

Page 9: ...INE will illuminate to show that mains power has been connected to the thermometer and the LED marked FAST will light when the batteries are low and the charger is charging at a high rate The charger...

Page 10: ...e shown SENS The following sensors may be configured Select SENS To configure for thermocouples select T C and the following screen will be displayed Select from one of the thermocouple options and pr...

Page 11: ...Enter the probe number and press OK if the probe number selected has already been configured then press Quit to return to the main measurement menu If you select a probe number that has no coefficien...

Page 12: ...tistical values of stored data log values Press Log Stats and the following values will be displayed Max and Min values Mean value Peak to peak value and Standard deviation The number of samples these...

Page 13: ...Press Clear to clear current value set and Edit to change the selected parameter To set the number of scan cycles press Edit and then key in the number of cycles required 1 to 9999 To set as continuo...

Page 14: ...ear on the display It will also be noticed that the flashing Busy blanks in the hold mode Pressing TRIG a second time returns the instrument to the run mode The second mode of operation may be set usi...

Page 15: ...and secondly from a calculation point of view the change in reading will be displayed without further calibration being required OPT Selects the various instrument configurations The following options...

Page 16: ...ture automatically applying the corrections The accuracy of the internal reference is approximately 0 3 C to achieve more accurate results an external reference may be used Ext When using an external...

Page 17: ...a diagnostic tool to ensure that there is no corruption of these constants Select the Cal option and you now enter the calibration routine The display will ask for a passcode to be entered the defaul...

Page 18: ...Ohms 1 Resistance Standard Type RS3 62 5 Value 62 5 Ohms 1 mV source 75 0 75 mV 1 Pt100 sensor and measuring instrument 1 Calibration Lead Type TS01 The Value of the resistance standards should be kn...

Page 19: ...to ensure the validity of all stored results and printed certificates Isothermal Technology Ltd accept no responsibility for any errors caused by inappropriate use incorrect set up or any other cause...

Page 20: ...ton Highlight PROGRAMS Select Isotech Select Calpad Note Ensure the the TTI 7PLUS is switched on and connected to the PC before starting the software Configuring for the TTI 7PLUS From File Config sel...

Page 21: ...ines Hardware handshaking for each character transfer uses the RTS and CTS lines RTS is an output from the instrument which indicates its receiver status When asserted low it indicates that it is read...

Page 22: ...re all incoming interface commands and the instrument will remain in LOCAL control mode The result of each temperature measurement is sent to the interface output with the resolution and units as per...

Page 23: ...ate command keywords but in this case the path remains at the same level This is not implemented for the instrument series The semicolon is treated as a non recognised character and the command error...

Page 24: ...contain ASCII characters which are placed between a pair of double quotation marks i e Output Data Format Non reading queries 80 ASCII character string Reading queries IEEE SDDDD DDD lf RS 232 SDDDD...

Page 25: ...do not trigger a measurement If the instrument is in the trigger run mode it will change to the trigger hold mode or if in the trigger single mode it will remain in the trigger single mode CONFigure C...

Page 26: ...lues are as per the tables above Does not affect the current measurement MEASure Group These work similarly to the CONFigure commands except that an immediate measurement is performed with the result...

Page 27: ...OLTage Returns the voltage produced by the thermocouple probe as displayed on the screen This is calculated from the voltage measured at the input terminals of the instrument plus the equivalent volta...

Page 28: ...lable for RS 232 only TRIGger Group TRIGger MODE SINGle INFinite Setting the mode to SINGle means that READ will take one reading and send it to the output buffer Setting the mode to INFinite means th...

Page 29: ...list timer cycle delay time slx SL1 SL2 SL3 SL4 Specify scan list Returns the delay time 00 00 00 to 99 59 59 SAMPle Group SAMPle COUNt SAMPle COUNt slx The SAMPle COUNt command queries the general t...

Page 30: ...ternal scanner cards slx SL1 SL2 SL3 SL4 scanlist A0 A1 A2 A3 A4 B0 B1 B2 B3 B4 To configure a scanning list specify the scanning list number followed by a list of channels that are required to be act...

Page 31: ...Tiate and FETCH are ignored DATAlogger MODE off on Turn the datalog mode off on When turned on triggering is controlled by STARt STOP and STEP and all readings are sent to the datalog memory It termin...

Page 32: ...m location 1 to POINts Sending ABORt will cancel the operation The output format is record no channel temperature unit date time DATAlogger POINts Queries the number of readings actually stored in the...

Page 33: ...on when this command was sent then the autozero function will be turned off Note that the MEASure commands affect channel sensor current mode etc and will therefore reset the mean and standard deviati...

Page 34: ...logged data format dd mm yy mm dd yy SYSTem DATE FORMat Queries the system calendar clock date format Returns DD MM YY or MM DD YY SYSTem DATE dd mm yy or mm dd yy Sets the system calendar clock date...

Page 35: ...Summary Byte is also cleared The Enable Register is a mask which allows any of the bits from the Event Register to set the appropriate bit in the Status Byte Setting an enable bit to 1 enables the ev...

Page 36: ...Reg Condition Event Enable Bit Register Reg Reg 0 0 0 0 Temperature Range erature Range 0 0 0 0 0 0 Under Lower Limit Over Upper Limit 0 0 0 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 0 0 0 0 Measuring 0 0...

Page 37: ...lected bits are then reported to the Status Byte STATus QUEStionable ENABle Queries the Questionable Data Enable Register The instrument returns a binary weighted decimal value representing the bits s...

Page 38: ...OPC Sets RQS on completion of the log Alternatively the application programme may continuously poll the Operation Complete bit using ESR Operation Registers Measuring bit 4 This bit is set when a new...

Page 39: ...etect when a measurement is complete and data is available Alternatively for IEEE the corresponding enable bit in the Service Request Enable Register can be set so that a Service Request SRQ interrupt...

Page 40: ...ing function The following command sequence can be used SRE 16 Enables the Message Available bit to trigger a service request CLS Clears the Message Available bit DATA START Initiates data logging OPC...

Page 41: ...pe rj mode ext rj standard TEMPerature RTD type standard con current mode root 2 CONFigure MEASure TEMPerature TC type rj mode ext rj standard TEMPerature RTD type standard con current mode root 2 CHA...

Page 42: ...ue all POINts Mathematical Operation Commands CALCulate AVERage MINimum AVERage MAXimum AVERage AVERage AVERage COUNt AVERage PEAK AVERage SDEV Autozero andp Rolling Mean Standard Deviation Commands S...

Page 43: ...T WAI Status Reporting Commands STATus QUEStionable CONDition QUEStionable EVENt QUEStionable ENABle enable value QUEStionable ENABle ESR ESE enable value ESE OPC STATus OPERation CONDition OPERation...

Page 44: ...Data Carrier Detect not connected 2 RXD Received Data input 3 TXD Transmitted Data output 4 DTR Data Terminal Ready not connected 5 GND Signal Ground 6 DSR Data Set Ready not connected 7 RTS Request T...

Reviews: