To select a scanning list pres SL1 to SL4, to turn off the scanning press Off. With the scanning list selected it is now
possible to edit the list press the Edt key
The channels included in the scanning sequence can now be selected by pressing the Add key or deleted with the Del
key; the highlighted channel will be added or deleted. To select a different channel use the < > keys NOTE: if
scanner cards are fitted then an additional 4 channels will be available per scanner card. Press OK to return to
previous screen.
From the SCAN main menu select TIMER to set the cycle and time intervals.
To select the parameter to change, use the < > keys and the selected parameter will flash. Press Clear to clear
current value set, and Edit to change the selected parameter.
To set the number of scan cycles press Edit and then key in the number of cycles required, 1 to 9999. To set as
continuous scan press Clear. Using the > key select Delay and Edit key in the required delay time between scans
hh:mm:ss this will also be the delay time before the first scan starts. Press OK to confirm set time and > to select the
reading rate. Press Edit and key in the reading rate hh:mm:ss this will be the time between each measurement.
Note: allow sufficient time for temperature to be measured and have regarding for stabilisation time of measured
sample. Press OK to confirm, and Quit to exit to main measurement screen
The thermometer is now set in the Hold mode and the scan sequence is started by pressing the TRIG key. A timer
will show in place of the Hold and will count down to the measurement scan start.