:LOADer
A valid command is formed by following the tree structure from a root node until a node
is reached with no further nodes below it, for example in the above command tree we
may use:
MILLiskaner:RJC:TEMPerature?
Keywords can be shortened to the first four letters (or 3 if the last letter is a vowel). To
indicate this, the notation uses upper-case to indicate required letters and lower-case to
indicate optional letters (NB: all commands are case-insensitive). For example, valid
forms of the above command include:
milliskanner:rjc:temperature?
mill:rjc:temp?
MILL:RJC:TEMP?
Mill:Rjc:Temperature?
To shorten the commands, default (optional) keywords are enclosed in square brackets
and may be omitted. For example, in the case of the command:
[ROUTe:]OPEN[:ALL]
Valid forms of this command include:
Route:Open:all
ROUT:OPEN:ALL
open