IFX-F100
User Manual
45
Flow_Transmitter_IFX-F100_EN_2018_March_Rev.0
9
Specification
General
Measuring principle : Ultrasonic time difference
correlation principle
Flow velocity range : 0.01 ... 25 m/s
Resolution : 0.25 mm/s
Repeatability : 0.15 % of measured value ± 0.015 m/s
Accuracy : ± 1 ... 3 % of measured value depending on application,
± 0.5 % of measured value with process calibration
Turn down ratio : 1/100
Gaseous and solid content of liquid media : < 10 % of volume
Flow transmitter
Enclosure : Wall or pipe mounted housing
Degree of protection : IP 66 according EN 60529
Operating temperature : -10 ... 60 °C (14 ... 140 °F)
Housing material : Die cast aluminium
Flow channels : 1
Power supply : 100 ... 240 V AC 50/60 Hz,
9 ... 36 V DC, specials upon request
Display : Optional LCD graphic display, 128 x 64 dots, backlit
Keypad : Optional four button internal keypad or programmer
Dimensions : H 120 x W 160 x D 80 mm without cable glands
Weight : Approx. 750g
Power consumption : < 5 W
Display damping : 0 ... 99 s
Measurement rate : 1Hz standard, higher rates on application
Response time : 1 s
Operating languages : Italian, English
Quantity and units of measurement
Volumetric flow rate : m3/h, m3/min, m3/s, l/h, l/min, l/s,
USgal/h (US gallons per hour),
USgal/min, USgal/s, bbl/d (barrels
per day), bbl/h, bbl/min, bbl/s
Flow velocity : m/s, ft/s, inch/s
Mass flow rate : g/s, t/h, kg/h, kg/min
Volume : m3, l, gal (US gallons), bbl
Mass : g, kg, t
Heat flow : W, kW, MW (only with heat quantity measurement option)
Heat quantity : J, kJ, MJ (only with heat quantity measurement option)
Sig dB (signal), noise dB, SNR,
C m/s (sound speed), CU (housing temperature)
Tin, Tout (inlet and outlet temperature)
Internal data logger
Storage capacity: In excess of 1 million data points (16MB)
Logging data: Up to ten selected variables
Communication
Serial interface : RS 232, RS 485, Modbus RTU, HART (optional)
Data : Measured data, parameter set and configuration,
Software
Functionality : Downloading of measured values/parameter sets, graphical presen-
tation, list format, export to third party software, on-line transfer of measured data
Operating systems : Windows 2000, NT, XP, Vista, 7; Linux; Mac (optional)