3.7.1
Sensor pipe mounting configurations
3.7.2
Acoustic coupling gel
In order to obtain acoustical contact between the pipe and the sensors, apply
a bead of acoustic coupling gel lengthwise down the centre of the contact
area of the sensors.
3.7.3
Correct positioning of the sensors
Always mount the transducer pair so that the free front edges of the sensors
face each other.
There is a different engraving on the top of each transducer. The transducers
are mounted correctly if the engravings on the two transducers form an ar-
row. The transducer cables should point in opposite directions.
Later, the arrow, in conjunction with the indicated measured value, will help
to determine the direction of flow.
Illustration 4: Sensor pipe mounting configurations
Illustration 5: Correct positioning of the sensors
Correct sensor
position