9.
MAINTENANCE
The
Staff in charge of the appliance
must control and respect the expiry dates for maintenance, given in the table
below, calling the authorised
Technical After-sales assistance
when indicated.
OPERATION
FREQUENCY
ORDINAR
Y
EXTRAORDINAR
Y
PERSONNEL
AUTHORISED
D
epending o
n the
Use an
d Necessit
y
Monthly
Ev
er
y 6
m
onths
An
nual
CLEANING THE EXTERNAL SURFACES
X
X
USER
CLEANING THE ACCESSIBLE INTERNAL PARTS
(without the use of tools)
X
X
CONTROL POWER SUPPLY CABLE,
PLUGS AND / OR ELECTRICAL SOCKETS
X
X
INTEGRITY CONTROL SEAL
X
X
FILTER CLEANING CONDENSING UNIT
(whenever present)
X
X
CLEANING THE DEFROSTING WATER COLLECTION TRAY
X
X
TECHNICAL
ASSISTANCE
CONDENSER CLEANING
X
X
X
CHECK COMPRESSOR OIL LEVEL
(whenever present)
X
AIR TANK DRAINING
(whenever present)
X
X
CONTROL PNEUMATIC CONNECTIONS
(whenever present)
X
X
INTEGRITY CONTROL PIPE COOLING SYSTEM
X
X
INSPECTION OF CABLES
INTERNAL CONNECTIONS AND POWER
X
X
CLEANING CONDENSATE DRYING SPONGES
(whenever present)
X
X
LAMP / LED REPLACEMENT
(whenever present)
X
CONTROL PANEL REPLACING
(electronic control unit - thermostat - etc.)
X
REPLACEMENT POWER SUPPLY CABLE, PLUGS AND / OR ELECTRICAL SOCKETS
X
ATTENTION
After all maintenance it is
mandatory
to perform all electrical safety tests in accordance with
the IEC EN 50106 Standard.