21
|
Electrical Wiring
IM-Raptor-1.3.6-EN
6.2 Connecting AC Power
For 100-240 VAC rated equipment, protection for earth fault is provided by max. 20 A branch circuit from AC input in
building installation. The protection in the building installation is relied upon for short-circuit backup protection.
a.The specification for AC breaker is 5 A, 2P, 277 VAC (min) Circuit breaker, Thermomagnetic or equivalent type.
b. The plug connector and wire gauge sizing is to be selected with appropriate design as per the Electrical code for a 60W,
1-phase device.
All equipment must be installed according to applicable local wiring codes and standards.
Always use cables that are rated for the operating ambient temperature of 85°C.
Note for IT power distribution systems:
1.This product is also designed for IT power distribution system with phase-to-phase voltage 230 V.
2.This equipment must be connected to an earthed mains socket-outlet.
Figure
A:
AC
Wiring
Diagram
L—stands for Live
N—stands for Neutral
NO—Normally Open (open = open circuit = not creating a path for the current)
NC—Normally Closed (closed = short circuit = creating a path for the current)
PE—Protective Earth (earth Ground point in the electrical circuit)