28
IRIDEX IQ 577
®
/IQ 532
®
Laser Systems Operator Manual
15510-EN Rev G
Event / Error Code
Error Message
Cause
User Action(s)
E06003
Missing Pulse error.
System unable to deliver laser
pulse when it was expected.
Check connections and
turn laser key OFF for
5 seconds then back ON.
W0001
Verify a 577 nm eye
safety filter is in place.
Confirmation of eye safety filter
is required before laser enters
Treat mode.
If using a 2-position filter,