26
IRIDEX IQ 577
®
/IQ 532
®
Laser Systems Operator Manual
15510-EN Rev G
Error Messages
System Errors
System errors display a message window (example below). When this screen is displayed, the system
has detected an interruption in one or more of the sub-systems.
User Action: Turn the keyswitch Off and then On. The system will attempt to correct itself. If the error
persists, write down the error code (example: E05002) and contact IRIDEX Service.
Error Code
Error Message
E05002
Emergency STOP pressed. Turn key off for 5 seconds then on.
E00701
System controller watchdog failure.
E01003, E01009
System needs calibration.
E03002, E03003
Invalid sensor reading.
E03010, E03020, E03040
Laser temperature invalid.
E03050
Heat sink reading invalid.
E04018, E04033, E04040,
E04050, E04051, E04052,
E04120, E04121, E04950, E04951
Voltage supply out of range.
E04099
Laser watchdog failure.
E06001, E06010
Laser power output out of range.
E06006, E06007
Photocell detector readings do not match.
E06030, E06102
Invalid laser output detected.
E06100
Photocell detector not responding.
E06101
Laser output detected in wrong port.
E06200, E06201
Invalid current detected at LCM shunt.
E08000
Software load failure in UIM.