EndoProbe and ENT Probe Operator Manual
13103F-EN
26
Concluding Patient Treatment
1. Place the console in Standby mode.
2. If desired, record the number of exposures and any other treatment
parameters.
3. Turn off the console and remove the key to prevent unauthorized use.
4. Disconnect the probe fiber-optic connector from the console
fiber-optic port. Cover the fiber-optic port with the protective cap.
5. Cut the probe in half with scissors so it cannot be used again. Discard
both probe pieces in accordance with your hospital’s procedures for
disposing of hazardous materials.
6. If using a SmartKey, remove it from the console and store it.
7. Disconnect the eye safety filter from the operating microscope and
store it.
8. If desired, clean and store the console.
9. If desired, remove the laser warning sign from the treatment room door.