Rated maximum output power
of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d= 1.2
P
80 MHz to 800 MHz
d= 1.2
P
800 MHz to 2.5 GHz
d= 2.3
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Declaration – electromagnetic emissions and immunity for EQUIPMENT and SYSTEMS that are not LIFESUPPORTING
and are specifi ed for use only in a shielded location.
Table 2 The iRenew™ Plus Pain Relief & Recovery System declaration – electromagnetic immunity
The iRenew™ Plus Pain Relief & Recovery System
is intended for use in the electromagnetic environment specifi ed below.
The customer or the user of the iRenew™ Plus Pain Relief & Recovery System should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Conducted RF IEC
61000-4-6
V/m
80 MHz to 2.5 GHz
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the EQUIPMENT or
SYSTEM including lead wires, than the recommended sep-
aration distance calculated from the equation applicable to
the frequency of the transmitter.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m