TROUBLESHOOTING
Always check the unit and accessories before use to prevent damage and defects; these are some of the simple
troubleshooting tips or you may call us at 855-723-2582.
If this happens…
Cause
Try this solution…
Device doesn’t turn on or
nothing visible on display.
No power. Battery power
is depleted.
• Ensure that the device has been fully charged.
• Press and hold ON/+ button for 3 seconds.
The battery indicated on
LCD display fl ashes
continuously.
Limited battery power.
• Ensure that the device has been fully charged.
The device does not seem
to be charging.
Limited battery power.
• Ensure the USB cable is properly connected in the
device and plugged into the AC adaptor and electrical
socket.
The device turns on, but
intensity cannot be in-
creased beyond “1” for
extended period.
Auto intensity reset
safety feature is initiated.
• Connect lead wires to device, electrodes to lead wires
and place on body part. 2 small or 1 large electrode
pad per channel is required.
• Replace used electrode pads. The quality of the gel
may be diminished.
The device turns on, but
does not generate electric
pulses.
Lead wires or electrode
pads are broken or
disconnected.
Treatment time expired.
• Replace/reconnect lead wires.
• Ensure lead wires are properly seated in CH1/CH2.
• Switch the device to the OFF position and then power
ON. Increase treatment time.