6.Regulatory Information
2450
2400-2570
Pulse modulation 217Hz, 28V/m
5240
5100-5800
Pulse modulation 217Hz, 9V/m
5500
5785
Input a.c. power port
Phenomenon
Basic EMC standard
Immunity test levels
Professional healthcare facility environment
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1 kV
Surges
Line-to-ground
IEC 61000-4-5
±0.5 kV, ±1 kV, ±2 kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands between 0.15MHz and
80MHz
80%AM at 1kHz
Voltage dips
IEC 61000-4-
11
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and
315º
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0º
Voltage
interruptions
IEC 61000-4-
11
0% UT; 250/300 cycles
Recommended separation distances between portable or mobile RF
communication device and detector:
Portable RF communications equipment, including antennas, can effect medical electrical
equipment. The warning should include a use distance such as “be used no closer than 30 cm (12
inches) to any part of the [ME EQUIPMENT or ME SYSTEM], including cables specified by the
manufacturer”.
Cable provided for EMC
Cable
Recommended
length
Shielded/Unshielded
Number
Cable
classification
AC power cable
1.8m
Unshielded
1 piece
AC power
DC power cable
3m
Shielded
1 piece
DC power
Ethernet cable
3.5m
Shielded
1 piece
Signal
Electromagnetic Compatibility (EMC)
Mars1417X requires special precautions regarding EMC and needs to be installed only by iRay
or authorized personnel and put into service according to EMC information provided in the user
manual. Mars1417X in use may be susceptible to electromagnetic interference from portable and
mobile RF communications such as mobile (cellular) telephones. Electromagnetic interference may
result in incorrect operation of the system and create a potentially unsafe situation.
Mars1417X conforms to this EN60601-1-2:2015 standard for both immunity and emissions.
71
User Manual of Mars1717X