Wireless Digital Flat Panel Detector Mars1417X
Generate Defect Template File
The process of generate defect map is quite similar with the one of gain map.
1.
On the “Create Defect” page, user can start the generating process by “Start” button.
And the process can be quit by “Cancel”.
2.
There are 8 images that need to be acquired, the UI gives the recommend KV and expected
image value, user should refer with them.
3.
If the option “Download to FPD after generation” is checked, the download UI will appear
after finishing generating the defect map which will takes a little time.
The field of “Index in FPD” means that the detector can store several correction maps and
choose one set to active as user wants.
The “Download files” part show the directory of the generated map stored on the workstation.
iRay Technology Co.