background image

9

Chapter 4

– Using the Control Panel

4.1.1  Description of Control Panel Keys

Power Key

The 

Power

key switches the IQAir system on and off.  When the system is switched off, the fan stops to run, but the

system will remain connected to the power supply (standby mode). The standby mode allows for automatic timer
start-up. In the standby mode, the different menu functions can be accessed. 

Tip: 

In a menu window,    the 

Power

key also serves as a quick exit key to return to the main display window.

Arrow (

p

) Key

When the system is switched on, the 

p

key allows the adjustment of the fan speed. In the enter mode, indicated by

the appearance of a black flashing cursor (see “Enter Key” below), the 

p

key is used to modify the selected setting in

the display window. Confirmed with the 

Enter

key, the enter mode is automatically terminated. The LCD will then

display the current menu settings for another 15 seconds before reverting to the main window display.

Chapter  4  –  Using the Control Panel

The IQAir system is operated and controlled via the electronic control panel which is located at the top of the front
locking arm.

Several operations can be completed via the electronic control panel:

Switching the system on and off

Controlling the fan speed and corresponding air delivery rate

Checking the remaining filter life of the individual filters

Setting the automatic timer

Resetting the Filter Life Monitor after replacing a filter

Locking the control panel to avoid tampering with the system’s settings

Choosing the desired display language

Setting the day & time

4.1 Description of Control Panel

LCD Display

The 2-line LCD displays important information about the system’s settings. 

In standby mode, (not running) without the timer activated, the first line
displays the IQAir model name and the word “Standby” is displayed on the
second line. If the timer is activated, the current day and time, and the next
programmed start and stop time will be displayed instead. 

If the IQAir system is running and the timer is not activated, the first line displays
the current speed setting and the second line displays the corresponding air
delivery rate of the system; however, when the timer is activated, the current
programmed start and stop times (Timer ON period) will be displayed on the
second line.

Main window: Standby mode

Main window: On mode

Menu window: Filter Life Monitor

Note: 

Airflow and remaining filter life

may be different on actual air cleaner.

LCD 

= Liquid

Crystal Display

Power

Key

Enter

Key

Menu

Key

Arrow

Key

HyperHEPA® Filter Life LED

Pre-Filter Life LED

Timer LED

Fan Speed LEDs

LED 

= Light Emitting Diode

Summary of Contents for Cleanroom H 13

Page 1: ...Cleanroom H13 User Manual Swiss Made...

Page 2: ...nces Replacement cords are available from IQAir Only use this air cleaner for its intended purpose of air cleaning in non industrial environments Do not use this air cleaner in areas with very high co...

Page 3: ...4 4 4 Day of the Week Timer 14 4 4 5 Timer ON OFF Fan Speed Selection Menu 14 4 4 6 Timer Information in the Control Panel 15 4 4 7 Time Day Setting 16 4 4 8 Filter Life Reset New Filter 16 4 4 9 Cha...

Page 4: ...to cope with your room size The actual hourly air volume cleaned by the system should be at least double the air volume of your room If the rate at which pollutants enter the room air is high or the i...

Page 5: ...tto u 4 2 m e t s y l o r t n o c e d n a t e k p o t e h t n re se e S e S C e d le o a T D D ay D e m i T Ti l a u n a m r e s u e h t o t r e f re og r p o t d e r i qu e r s i k c o cl e h t g in...

Page 6: ...ntrally the system is positioned the better the performance Position the system close to a power outlet and make sure that the power cord is positioned in a way that it does not present an obstacle in...

Page 7: ...gn features air intake openings on two sides The Pre Filter module consists of frame 1 and the Pre Filter which is secured inside the frame by four filter clamps The fan assembly comprises two housing...

Page 8: ...culate air HyperHEPA filter media 5 Low turbulence low velocity air is returned to the room via the EvenFlow diffuser 8 Chapter 3 The IQAir System s Components 3 2 Description of Removable Components...

Page 9: ...and off Controlling the fan speed and corresponding air delivery rate Checking the remaining filter life of the individual filters Setting the automatic timer Resetting the Filter Life Monitor after...

Page 10: ...or a reduction in filter efficiency gas phase filter is saturated Fan Speed LEDs These LEDs simulate the fan speed through the frequency of their rotation The faster the rotation of the fan speed LED...

Page 11: ...et per minute x 60 air changes per hour ACH air volume of the room cu ft L x W x H Note The air delivery rates of the IQAir system at the various fan speeds settings are listed on the Technical Specif...

Page 12: ...lter is beyond its useful life a negative number will be displayed How the Filter Life Monitor Works The basis for calculating the remaining filter life is the already elapsed operation time at the se...

Page 13: ...rsor starts to flash 3 Press the p key one time to activate the timer Note If the START and STOP times are identical the timer cannot be activated Note The time is displayed in a 24 hrs clock military...

Page 14: ...n and exits the enter mode 4 4 4 Day of the Week Timer The Day of the Week Timer allows the deactivation of the Timer ON period on certain days of the week In its default setting the Timer ON period i...

Page 15: ...ng The first line shows the fan speed and the second line shows the airflow The Timer LED is off System on Standby The first line shows the current day and time and the second line shows the next time...

Page 16: ...ter Life LED for the appropriate filter is currently flashing red 4 Press the Enter key to confirm that the selected filter has been replaced 5 To re confirm the filter change press the Enter key To e...

Page 17: ...er mode 17 Chapter 4 Using the Control Panel 4 4 9 Changing the Airflow Units 1 You can reach the Airflow Units Menu by pressing the Menu key eight times 2 You can change the air volume units which ar...

Page 18: ...en set in the Weekly Auto Timer Menu it is possible to switch the timer on and off with the remote control To switch the timer on press the Timer Modification key and then the Timer ON key To switch t...

Page 19: ...r of minutes The illustration of the IQAir system on the control panel shows the positions of the individual filters These positions correspond to the actual positions of the filters within the system...

Page 20: ...s and push the locking arms inwards simultaneously until they snap back into the housing top of the diffuser IQ Air 20 Chapter 6 Replacing Filters Opening Closing 6 3 1 Troubleshooting When Closing th...

Page 21: ...me from the filter 10 Make sure that the filter is inserted all the way into the frame Insert the clamps into the sliding grooves inside the frame and push them down securing the filter into place 6 5...

Page 22: ...r at www iqair com support where you can find answers to most frequently asked questions FAQs or contact us with your inquiry To expedite your service request please have the following information rea...

Page 23: ...ng fixture you will also need a medium size flathead screwdriver How to install the power cord fixture 1 Disconnect the power cord from the wall outlet Plug the connector of the power cord into the ai...

Page 24: ...ng the supplied tool 7 8 6 9 Plug the red connector of the ribbon cable back into the control panel Ensure that it is pushed in all the way 10 Tuck the ribbon cable underneath and reinsert the control...

Page 25: ...of positive pressure if placed within the room and negative pressure if placed outside the room OutFlow Complete ducting adapter kit for the creation of negative pressure if placed within the room an...

Page 26: ...ly usage of 10h on speed 3 U S FDA registered as a Class II Medical Device Registration No 2087558 Performance 99 97 for particles 0 3 microns indiv tested at highest fan speed 99 5 at 0 003 microns y...

Page 27: ...27 Notes...

Page 28: ...www iqair com Swiss Made IQ_MA_CRH13_NE_US_130724...

Reviews: