FALCO Instrument User Manual V1.2R
Unrivalled Gas Detect
ion.
Page 7 of 46
ionscience.com
Protect the PID sensor from exposure to silicone vapours as this may foul the
windows of the lamps and reduce response to some gases. This can usually be
remedied by polishing the lamp window with alumina powder.
Do not use abrasive or chemical detergents to clean the Falco instrument as this
may reduce the antistatic properties of the materials used, clean it using a damp
cloth only.
The Falco must not be exposed to atmospheres known to have an adverse effect
on Thermoplastic Elastomers or Polycarbonate.
Outside the items covered in this manual, the Falco must be serviced in a Non-
Hazardous environment and by ION Science Ltd authorised service centres only.
Substitution of components may impair intrinsic safety.
Ingress Protection: Continuous exposure to wet weather conditions should be
limited to less than one day and harsh water spray conditions should be avoided.
Proper Use: If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
The following Warnings, Cautions and Information notifications appear later in this manual where they are
applicable.
IF AN ALARM STATE IS TRIGGERED THE USER SHOULD LEAVE THE HAZARDOUS
ENVIRONMENT AND ACT IN ACCORDANCE WITH NATIONAL SAFETY
REGULATIONS.
THE CLEANING COMPOUND CONTAINS ALUMINIUM OXIDE AS A VERY FINE
POWER. THIS MAY CAUSE IRRITATION OF RESPIRATORY TRACT AND EYES.
(CAS Number 1344-28-1).
Internal components must be handled with clean hands and clean tools. The lamp
is fragile. Handle with great care. Never touch the window and do not drop.
Never refit a damaged Lamp.
The instrument MUST be re-calibrated after fitting a replacement or cleaned
Lamp.
The Falco has been designed to be used in hazardous environments