ionscience.com
Display Screen
OLED 64x128 pixels
WEEE Compliant
Compliant with the European Waste Electrical and Electronic Equipment
directive
Certification
II 2G Ex db ib IIC T4 Gb
Class 1, Div 1 Groups ABCD T4
ISO9001:2015
All specifications quoted are at calibration point and under the same ambient conditions.
Specifications are based on isobutylene calibration at 20 °C and 1000 mBar.
Un-Packing
All equipment shipped by Ion Science Ltd is packed in containers with shock absorbing filling to protect them
against physical damage.
Remove the contents carefully and check them against the packing list. Report discrepancies between the
contents and the packing list to Ion Science Ltd. Ion Science will not be responsible for discrepancies not
reported within ten days of your receipt of the shipment.
Every FALCO (new units and those returned from a Service Centre) must have a Certificate of Calibration
before you install it.
System Description
Outputs and Communications
FALCO has six communication outputs:
•
The on-board LCD and LEDs on the faceplate.
•
4 - 20 mA Current Loop.
•
RS485 Modbus.
•
Two programmable relays.
Real-time information is displayed on the LCD and transmitted on the 4 - 20 mA and RS485 channels.
You can program two alarms to operate at a chosen concentration of gas. The alarms will display a message on
the LEDs, energise the relays, and transmit a signal on the 4 - 20 mA channel.
The alarms and relays are individually programmable to the settings required by the site policy. You can
choose either alarm to energise either relay.
Both relays can be programmed to switch 60 VDC / 2 A or 50 VAC / 2 A maximum load.