background image

Page 7

7.3

LOW FLOW ERROR:

The flow through the sensor is controlled by the vacuum pump. The level of vacuum
required to maintain the required flow can be seen in the lower LH screen. Should the
sample intake of the sensor be partially or totally blocked, the instrument will first try to
compensate by increasing the vacuum. If unsuccessful, an error is reported.
You have the choice to confirm one of the following instructions: TRY TO CONTINUE
Instrument will try to resume normal operation. This is useful if the blockage can be
overcome with a higher vacuum. The LOW FLOW ERROR may also occur if the sensor
cap was changed for the capillary probe, this needs higher vacuum to maintain the
sample flow. 

BEFORE changing probes and sensors ALWAYS switch OFF

.

The ideal vacuum levels for a clean filter are: -100 to -300mb.

To fit the capillary probe, FIRST switch the instrument off. This allows the new flow levels
to be automatically compensated.

7.4

FLARE STALL:

This relates to sensor condition. Two reasons are possible for this.
1

Very high concentration of SF6 many times above the measuring range.

2

High vacuum in the probe caused by a partially blocked or aged sensor.

The instrument will attempt to compensate for these until a level is reached where the
detector will stall. At this point the instrument must be switched OFF and the sensor
changed.

7.5

LOW BATTERY VOLTAGE:

If the battery becomes discharged, the operator is warned of the pending shut down.

If main’s power cannot be made available, operation can be resumed for a short time, but
the instrument may now shut down at any time without further warning.

7.5

SWITCHING OFF:

During system shutdown, the processor updates the memory, releases vacuum, and
turns OFF the touch-screen. To shutdown, press one of the on/off keys for several
seconds. A time bar on the screen visualises the hold time.

When the time bar is complete, the button should be released. If it was held down longer,
the shutdown sequence will not finish properly.

8.0

SWITCHING OFF IN CASE OF FAULT:

Normally, shutdown is controlled as described above. Should this process be changed by
a serious failure, the instrument can still be switched off under hardware control. This
activates whenever one of the on/off keys is continuously held down for approx. 10
seconds.

Summary of Contents for GAS-CHECK SF6

Page 1: ...INSTRUCTION MANUAL GAS CHECK SF6 AN INSTRUMENT FOR THE LOCATION AND MEASUREMENT OF LEAKS IN SF6 FILLED ELECTRICAL SWITCHGEAR User Manual iss 01 Software Version V1 39BB...

Page 2: ...ion 6 7 3 Low flow error 7 7 4 Flare stall 7 7 3 Low battery voltage 7 7 5 Switching OFF 7 8 0 Switching OFF in case of fault 7 8 1 User setup 8 8 2 Set volume Contrast 8 9 0 Open Operator Level 8 9 1...

Page 3: ...p after fuse has been removed 15 16 3 Pre set passwords on delivery 15 16 4 Settings on despatch 16 16 5 Citizen printer 16 16 6 Spares and consummables 16 3 0 GENERAL This manual is addressed to thos...

Page 4: ...contain a microchip programmed with specific sensor information At Switch ON the console automatically reads sensor information Smart Sensors are not radioactive and contain nothing harmful Life is ap...

Page 5: ...may confidently measure This is OFF whenever the instrument is in standby or programming mode error condition or the actual reading is above 20 of the alarm level Red 2 NOT READY This light is on whe...

Page 6: ...used at any time During the short zero time the lights change from RED to GREEN Do not leak test during this time During normal use it is recommended to press the CIRCLE key occasionally to zero the d...

Page 7: ...for you press Exit Should this not be the case please refer to paragraphs Possible Fault Conditions During Operation and Switching Off in Case of Fault in this manual 7 1 MEASURING OPERATION The GAS C...

Page 8: ...entration of SF6 many times above the measuring range 2 High vacuum in the probe caused by a partially blocked or aged sensor The instrument will attempt to compensate for these until a level is reach...

Page 9: ...the present value of a parameter will already be entered You may then confirm or delete and re enter 8 2 Set volume Contrast Using the three slider type controls on this screen select contrast for be...

Page 10: ...vel alerts the ALARM LED in the handgun Only the alarm level corresponding to the chosen mode ml sec or ppm can be entered Valid alarm levels are for leak rates 0 5 E 6 5 E 7 ml s to 20 E 6 200 E 7 ml...

Page 11: ...be assigned for 6 different operators please refer to Changing Passwords section 10 4 10 3 Access codes 10 4 CHANGING PASSWORDS Use one of the eight buttons to select the password you wish to change t...

Page 12: ...Air is measured first which takes 4 seconds A time bar is displayed along with an audible bleep The display then instructs the operator to put the probe to the reference leak When this is done the fi...

Page 13: ...when the STORE key is pressed Pressing either the STORE or CLEAR buttons resets this 12 3 Display units The touch screen display and alarm level entries as well as reference leak values will operate...

Page 14: ...e following will occur The touch screen screen contrast sensor and sample flow regulation are switched off Vacuum is released Lifetime of all major components is saved If POWER DOWN is selected the in...

Page 15: ...r operation The main function of the filter is to block out short wave noise Filter Count represents the number of measuring cycles over which the signal must be continuously above or below the averag...

Page 16: ...ew sensor and sets these values in the console memory this may take 10 50 seconds 15 2 Changing the Probe To change the sensor You Must Power Down Never unplug the probe while the red lights on the co...

Page 17: ...er set up ALL 4 SWITCHES MUST BE IN THE OFF POSITION Switch the printer ON using the side mounted slide switch The POWER and SEL green lights should be lit Press SEL to turn OFF the light and use LF t...

Reviews: