11.0

Calibration

The GAS CHECK SF6 is unique in that completely separate values are stored for each of
the display modes. This means that if you want to measure ppm or gm/yr these values
must be calibrated before use. Calibration follows the same easy routine as with cc/sec
calibration using the 

Portable Reference Leak

. Regular calibration further improves the

precision of measurement over the standard (factory set) calibration.

For cc/sec (volume loss over time mode) You will need the Portable Reference Leak
(PRL), supplied as an option with the instrument and available from Ion Science Ltd.

The leak rate will be approximately 5x10-6cc/sec (5 E-6cc/sec)

Press Calibration. The screen now asks you to enter (or confirm) the leak rate of the PRL
(or gas concentration when in ppm mode). If this is to be changed, then use the keypad
to enter the SAME leak rate as printed on the PRL.

As long as the same PRL / gas concentration is used, this value needs to be entered only
once

Just press RETURN to confirm it each time you make a calibration.

On pressing 

RETURN 

the calibration process begins.

Background Air is measured first, which takes 4 seconds. A time bar is displayed along
with an audible ‘bleep’. The display then instructs the operator to put the probe to the
reference leak. When this is done, the final calibration step begins automatically and
takes 6 seconds.

During this time, it is important to hold the probe accurately in the reference leak,
otherwise an incorrect calibration may occur.

Should the leak be missed or the probe removed by mistake, just move the probe off the
leak. The time bar is then reset, and the process can be re started.

When calibration is complete, the instrument returns to normal mode. Now check the
portable reference leak, which should be the same reading (+/-5%). Repeat the
calibration if necessary, and remember to accurately hold the probe in the leak.

The Exit  button can be used to abort calibration at any time. The original calibration
factor then remains unchanged.

11.1

Set clock

The clock keeps time even while the instrument is off. Adjustment may be necessary for
the following: Summer / winter time changeover, Different time zone
Normal long term adjustment.
If the battery has been disconnected. 

Should this happen, the instrument automatically

opens the set date/time screen after power on.

This is quite straightforward; use the -/+ pads to change and when correct press Exit.

Summary of Contents for GAS-CHECK SF6

Page 1: ...INSTRUCTION MANUAL GAS CHECK SF6 AN INSTRUMENT FOR THE LOCATION AND MEASUREMENT OF LEAKS IN SF6 FILLED ELECTRICAL SWITCHGEAR User Manual iss 01 Software Version V1 39BB...

Page 2: ...ion 6 7 3 Low flow error 7 7 4 Flare stall 7 7 3 Low battery voltage 7 7 5 Switching OFF 7 8 0 Switching OFF in case of fault 7 8 1 User setup 8 8 2 Set volume Contrast 8 9 0 Open Operator Level 8 9 1...

Page 3: ...p after fuse has been removed 15 16 3 Pre set passwords on delivery 15 16 4 Settings on despatch 16 16 5 Citizen printer 16 16 6 Spares and consummables 16 3 0 GENERAL This manual is addressed to thos...

Page 4: ...contain a microchip programmed with specific sensor information At Switch ON the console automatically reads sensor information Smart Sensors are not radioactive and contain nothing harmful Life is ap...

Page 5: ...may confidently measure This is OFF whenever the instrument is in standby or programming mode error condition or the actual reading is above 20 of the alarm level Red 2 NOT READY This light is on whe...

Page 6: ...used at any time During the short zero time the lights change from RED to GREEN Do not leak test during this time During normal use it is recommended to press the CIRCLE key occasionally to zero the d...

Page 7: ...for you press Exit Should this not be the case please refer to paragraphs Possible Fault Conditions During Operation and Switching Off in Case of Fault in this manual 7 1 MEASURING OPERATION The GAS C...

Page 8: ...entration of SF6 many times above the measuring range 2 High vacuum in the probe caused by a partially blocked or aged sensor The instrument will attempt to compensate for these until a level is reach...

Page 9: ...the present value of a parameter will already be entered You may then confirm or delete and re enter 8 2 Set volume Contrast Using the three slider type controls on this screen select contrast for be...

Page 10: ...vel alerts the ALARM LED in the handgun Only the alarm level corresponding to the chosen mode ml sec or ppm can be entered Valid alarm levels are for leak rates 0 5 E 6 5 E 7 ml s to 20 E 6 200 E 7 ml...

Page 11: ...be assigned for 6 different operators please refer to Changing Passwords section 10 4 10 3 Access codes 10 4 CHANGING PASSWORDS Use one of the eight buttons to select the password you wish to change t...

Page 12: ...Air is measured first which takes 4 seconds A time bar is displayed along with an audible bleep The display then instructs the operator to put the probe to the reference leak When this is done the fi...

Page 13: ...when the STORE key is pressed Pressing either the STORE or CLEAR buttons resets this 12 3 Display units The touch screen display and alarm level entries as well as reference leak values will operate...

Page 14: ...e following will occur The touch screen screen contrast sensor and sample flow regulation are switched off Vacuum is released Lifetime of all major components is saved If POWER DOWN is selected the in...

Page 15: ...r operation The main function of the filter is to block out short wave noise Filter Count represents the number of measuring cycles over which the signal must be continuously above or below the averag...

Page 16: ...ew sensor and sets these values in the console memory this may take 10 50 seconds 15 2 Changing the Probe To change the sensor You Must Power Down Never unplug the probe while the red lights on the co...

Page 17: ...er set up ALL 4 SWITCHES MUST BE IN THE OFF POSITION Switch the printer ON using the side mounted slide switch The POWER and SEL green lights should be lit Press SEL to turn OFF the light and use LF t...

Reviews: