background image

Page 12

12.0

User set-up

This gives access to a number of parameters which determine the behaviour of the
instrument. They are described as follows and all are selected using the up/down arrows
and changed using the -/+ buttons.

12.1

Peakhold (gauge).  This due to the short recovery time after detecting a leak, this

function is helpful. The peak value will be held on the handgun display for the time you
select. In position OFF, no peak hold occurs.

12.2

Peakhold (screen) on the touch-screen, two leak rate displays are visible: the

main display in big digits, and the auxiliary display below it. One of them always shows
the ‘live’ reading, while the other holds a peak value (if selected). This is factory pre set
to AUX DISPLAY.

In this position (AUX DISP) the big display shows the live reading, while the auxiliary
display holds the peak. If set to MAIN DISPLAY, the peak hold is held on the big display,
while the auxiliary display shows the live reading.

The peak hold seen on the console display represents the value that is stored in memory
when the STORE key is pressed. Pressing either the STORE or CLEAR buttons resets
this.

12.3

Display units

The touch-screen display and alarm level entries, as well as reference leak values, will
operate in the units you select.

Use the +/- buttons to change between E-6 ml/sec, E-7 ml/sec, gm/yr and ppm.

For gm/yr and ppm measurements, all the above parameters are held separately and 

do

not need to be re-entered

 when changing between leak rate and concentration modes.

There is also a separate calibration factor for each mode.

Gm/yr (weight loss) and ppm (concentration) measurement should not be used for leak
detection. Leaks are measured in volume loss/time (ml/sec). .ppm mode demands a
higher sensitivity, than leak rate mode. When measuring low ppm concentrations the
display responds considerably slower. Response time depends on the actual
concentration and, at very low levels, may exceed 10 seconds.

In the present software version, the ppm mode is
still in an experimental state. Therefore all
information is approximate. For precise ppm
measurement it is recommended that calibration
is made to a 10ppm SF6 in air source. For more
information please contact ISL.

Desens Factors. Allow large leaks or large concentrations to be displayed.

Summary of Contents for GAS-CHECK SF6

Page 1: ...INSTRUCTION MANUAL GAS CHECK SF6 AN INSTRUMENT FOR THE LOCATION AND MEASUREMENT OF LEAKS IN SF6 FILLED ELECTRICAL SWITCHGEAR User Manual iss 01 Software Version V1 39BB...

Page 2: ...ion 6 7 3 Low flow error 7 7 4 Flare stall 7 7 3 Low battery voltage 7 7 5 Switching OFF 7 8 0 Switching OFF in case of fault 7 8 1 User setup 8 8 2 Set volume Contrast 8 9 0 Open Operator Level 8 9 1...

Page 3: ...p after fuse has been removed 15 16 3 Pre set passwords on delivery 15 16 4 Settings on despatch 16 16 5 Citizen printer 16 16 6 Spares and consummables 16 3 0 GENERAL This manual is addressed to thos...

Page 4: ...contain a microchip programmed with specific sensor information At Switch ON the console automatically reads sensor information Smart Sensors are not radioactive and contain nothing harmful Life is ap...

Page 5: ...may confidently measure This is OFF whenever the instrument is in standby or programming mode error condition or the actual reading is above 20 of the alarm level Red 2 NOT READY This light is on whe...

Page 6: ...used at any time During the short zero time the lights change from RED to GREEN Do not leak test during this time During normal use it is recommended to press the CIRCLE key occasionally to zero the d...

Page 7: ...for you press Exit Should this not be the case please refer to paragraphs Possible Fault Conditions During Operation and Switching Off in Case of Fault in this manual 7 1 MEASURING OPERATION The GAS C...

Page 8: ...entration of SF6 many times above the measuring range 2 High vacuum in the probe caused by a partially blocked or aged sensor The instrument will attempt to compensate for these until a level is reach...

Page 9: ...the present value of a parameter will already be entered You may then confirm or delete and re enter 8 2 Set volume Contrast Using the three slider type controls on this screen select contrast for be...

Page 10: ...vel alerts the ALARM LED in the handgun Only the alarm level corresponding to the chosen mode ml sec or ppm can be entered Valid alarm levels are for leak rates 0 5 E 6 5 E 7 ml s to 20 E 6 200 E 7 ml...

Page 11: ...be assigned for 6 different operators please refer to Changing Passwords section 10 4 10 3 Access codes 10 4 CHANGING PASSWORDS Use one of the eight buttons to select the password you wish to change t...

Page 12: ...Air is measured first which takes 4 seconds A time bar is displayed along with an audible bleep The display then instructs the operator to put the probe to the reference leak When this is done the fi...

Page 13: ...when the STORE key is pressed Pressing either the STORE or CLEAR buttons resets this 12 3 Display units The touch screen display and alarm level entries as well as reference leak values will operate...

Page 14: ...e following will occur The touch screen screen contrast sensor and sample flow regulation are switched off Vacuum is released Lifetime of all major components is saved If POWER DOWN is selected the in...

Page 15: ...r operation The main function of the filter is to block out short wave noise Filter Count represents the number of measuring cycles over which the signal must be continuously above or below the averag...

Page 16: ...ew sensor and sets these values in the console memory this may take 10 50 seconds 15 2 Changing the Probe To change the sensor You Must Power Down Never unplug the probe while the red lights on the co...

Page 17: ...er set up ALL 4 SWITCHES MUST BE IN THE OFF POSITION Switch the printer ON using the side mounted slide switch The POWER and SEL green lights should be lit Press SEL to turn OFF the light and use LF t...

Reviews: