background image

Intra-Automation Technical 

Information 

IA-N24 

 

 

 

- 4 - 

 

3 Technical details 

 

Inputs 

Type Measuring 

ranges 

Parameters 

Overloads 

Errors 

-11 mV…-10 mV…60 mV…66 mV 
66 mV…-60 mV…60 mV…66 mV 
-0,5V…0 V…10 V…11 V 
-11 V…-10 V…10 V…11 V 

Input resistance: 

 1 M

 

-1 mA…0 mA…20 mA…22 mA 

Input resistance: 10 

 

 1% 

IA-N24S 

3,6 mA…4 mA…20 mA…22 mA 

Input resistance: 10 

 

 1% 

Short term overload (1s): 
- voltage input: 10 Un 
- current input: 5 In 
Sustained overload: 
110 % Un, 110 % In 

Basic error: 

 (0,2% of range + 1 digit) 

Additional error

 from ambient 

temperature changes:  

(50 % of basic error)

 

Pt100 -50…150 

°C 

 -50…400 

°C 

Current flowing though the 
sensor: 

 300 µA 

Resistance of wires conn. 
RTD with the meter: 
- max. 5 

 (per wire) for 

automatic compensation 
- max. 10 

 (per wire) for 

manual compensation 

Thermocouple 
Type J 

-50…1200 °C 

IA-N24T 

Thermocouple 
Type K 

-50…1370 °C 

 

Short term overload (1s): 
Input of sensors: 30 V 

Basic error: 

 (0,2% of range + 1 digit) 

Additional errors:

  

- compensation of cold junction 
temperature changes: 

 0,2 % of range 

from ambient temperature 
changes:  

(50 % of basic error)

 

1…100…120 V AC 
2,5…250…300 V AC 
4…400…600 V AC 
20…500 V AC 
(in voltage range: 24…480 V) 

Input resistance: 

 2 M

 

0,01…1…1,2 A AC 

Input resistance: 10m

 

10% 

IA-N24Z 

0,05…5…6 A AC 

Input resistance: 2m

 

10% 

Short term overload (1s): 
Voltage input: 2 Un (

1000 V), 

Current input: 10 In 
Sustained overload 
150 % Un (for 400 V input) 
120 % (for remaining inputs) 
120 % ln 

Basic error: 

- voltage and current: 

 (0,5% of range + 1 digit) 

- frequency: 

 (0,02% of range + 1 digit) 

Additional error

 from ambient 

temperature changes:  

(50 % of basic error) 

IA-N24H 

0…100…120 V DC 

 

0…250…275 V DC 

 

-120…-100…100…120 VDC 

 

-300…-250…250…300 VDC 

 

-600…-400…400…600 VDC 

Input resistance: 

 2 M

 

 

-1,2…-1…1…1,2 A DC 

Input resistance: 10m

 

10% 

 

-6…-5…5…6 A DC 

Input resistance: 2m

 

10% 

Short term overload (1s): 
Voltage input: 2 Un (

1000 V), 

Current input: 10 In 
Sustained overload 
150 % Un (for 400 V input) 
120 % (for remaining inputs) 
120 % ln 

Basic error: 

 (0,2% of range + 1 digit) 

Additional error

 from ambient 

temperature changes:  

(50 % of basic error)

 

 

Outputs 

For N24S and N24T 

Output for supply of external transducers 

24 V 

 5 %, 30 mA 

 

External features 

Weight 

 0,25 kg 

 

Overall dimensions 

96 x 48 x 64 mm (with terminals) 

 

Protection grade 

IP65 (Front side) 

IP10 (Terminal side) 

Display 

4-digit LED display, 20 mm height, red 

Indication range: -1999…9999 

 

Rated operating conditions 

Supply voltage 

230 V 

 10 % AC (45…65 Hz); 110 V 

 10 % AC (45…65 Hz); 

24 V 

 10 % AC (45…65 Hz); 

85..253 V AC (40…400 Hz)or DC; 20..40 V AC (40…400 Hz)or DC; 

Power consumption 

 6VA 

Temperature Ambient: 

-10…23…55 °C 

Storage: -25…85 °C 

Relative Humidity 

 95 % 

Condensation inadmissible 

Operating position 

Any 

 

Preheating time 

30 min 

 

Averaging time 

 0.5 s 

1 second default set 

 

Safety and compatibility conditions 

Electromagnetic compatibility 

Noise immunity 

acc. to EN 61000-6-2 

 

Noise emissions 

acc. to EN 61000-6-4 

Isolation between circuits 

Basic 

Pollution grade 

Installation category 

III (for the 400 V option – category II) 

Maximal phase-to-earth operating 
voltage 

For supply circuits: 300 V, for measuring circuits: 
600 V, cat. II 

 

For other circuits: 50 V 

Altitude above sea level 

 2000 m 

acc. to EN 61010-1 

 

Summary of Contents for IA-N248

Page 1: ...THE EXPERT IN LEVEL AND FLOW Accessories Digital Panel Meter IntraDigit Series IA N24 Technical Information 02 2011...

Page 2: ...cal Information Intra Automation 1 Intra Automation Technical Information 02 2011 Technical details subject to be changed without notice For comments regarding this brochure please contact info intra...

Page 3: ...ation Technical Information IA N24 2 IA N24 IntraDigit Digital Panel Meter List of Contents Chapt Title Page 1 Features 3 2 Application Example 3 3 Technical details 4 4 Electrical Connection 4 5 Orde...

Page 4: ...grammer Precision of displayed results decimal point measurement averaging time recounting of indications individual characteristic automatic or manual compensation temperature of cold ends for J K or...

Page 5: ...f range 1 digit Additional error from ambient temperature changes 50 of basic error IA N24H 0 100 120 V DC 0 250 275 V DC 120 100 100 120 VDC 300 250 250 300 VDC 600 400 400 600 VDC Input resistance 2...

Page 6: ...nnection of the N24Z and N24H meters for voltage measurement frequency only N24Z Electrical connection of the N24Z and N24H meters for current measurement Connections of N24T measuring inputs Resistan...

Page 7: ...e XX custom made Language E English X other Certification 0 without additional quality inspection certificate 1 with additional quality inspection certificate X On request IA N24 only for N24S and N24...

Page 8: ...int 000 0 for I U Averaging time 1 s Upper overflow of measurement 9999 Lower overflow of measurement 1999 Individual characteristic enabled Parameter a of the individual characteristic 5 Parameter b...

Page 9: ...for industrial measurement tasks For more information please contact us contact details on the backside of this brochure Flow measurement Itabar Flow Sensor IntraSonic IS210 Ultrasonic Flow Meter Lev...

Page 10: ...BENELUX Intra Automation GmbH B V Intra Automation HTP Otto Hahn Str 20 PO Box 10 41515 Grevenbroich 4730 AA Oudenbosch GERMANY THE NETHERLANDS 49 0 21 81 7 56 65 0 31 0 165 32 22 01 49 0 21 81 6 44...

Reviews: