Simpli Series System Instructions for Use
30
01-5034-00 R1
Appendix B
–
Interson Probe System Specifications
Imaging Mode
B Scan
Functions
•
Standard USB port (2.0) connectivity
•
Multiple freeze methods: button on
probe, keyboard, or button on screen
•
Auto image save on Freeze
Image Resolution
•
0.1 to 2.0 mm resolution*
Gray Shades
•
True 256 (8 bits) shades of gray
Sector Size
•
Rectangular* or 60° sector*
Transducers
•
High bandwidth, multiple elements:
5.0 MHz and 7.5 MHz nominal
frequencies*
Depth Selections
•
2 cm through 24 cm depths*
Measurements
•
Distance, angle, area, circumference, and
volume measurements
Signal Processing
•
TGC controls: near, mid, and far
•
Contrast and image intensity controls
•
Frame averaging
•
Interpolation
Archives Functions
•
Exam data
•
Cine buffer range 32-1024 frames
•
Open system architecture
Power Supply Requirements
•
DC 5.0 VDC, ±5% at 500 mA (max) 2.5
watts (max) obtained from the USB 2.0
port.
Environmental
•
Max operating temperature: 50°C (122°F)
•
Min operating temperature: -10°C (14°F)
•
Operating humidity range: 20-80%
Non-condensing
Storage Temperature
•
-10°C to 50°C (14°F to 122°F)
*Probe dependent