Simpli Series System Instructions for Use
11
01-5034-00 R1
2.
Check the expiration date on the solution that is being used.
3.
Use only solutions that are within the expiration date.
Y
WARNING
The level of disinfection required for a device is dictated by the type of tissue it will contact during use. To
avoid infection, ensure the disinfectant type is appropriate for the equipment. For information, see the
disinfectant label instructions and the recommendations of the Association of Professional in Infection
Control and Epidemiology (APIC) and the U.S. Food and Drug Administration (FDA).
Y
WARNING
Using a non-recommended disinfection solution, incorrect solution strength, or immersing a probe deeper
or for a period longer than recommended by disinfectant manufacturer can damage or discolor the probe
and will void the probe warranty.
For more information on probe warranty please refer to Appendix L.
Y
WARNING
Disinfect probes using only liquid solutions. Using autoclave, gas (EtO), heat or radiation to sterilize or
other non-Interson-approved methods will permanently damage the probe and void the warranty.
1.
Examine the probe for damage such as cracks, splitting, sharp edges or projections. If damage is
evident, abandon disinfection, discontinue use of the probe, and contact a customer service
representative.
2.
Mix the disinfecting solution that is compatible with the probe according to label instructions for
solution strength. A disinfectant qualified by the FDA 510(k) process is recommended.
3.
Immerse the probe up to the cable in the disinfecting solution for the recommended length of
time per the manufacturer. Do not immerse any part of the probe’s cable or cable con
nector.
4.
Using the instructions on the disinfectant or sterilization label, rinse the probe up to the strain
relief, and then air dry or towel dry with a clean cloth.
Training
This USB probe system is intended to be used by trained medical professionals only.
The specific probe functions are described in this manual and are also available through the SimpliVue
system software Help Menu.
Storage
When the probe is not being used, it should be stored in a clean, dry area.
Y
CAUTION
Disinfect the probe before use to avoid it becoming a source of infection.
To prevent damage to the probe, do not store in areas where it might be exposed to:
•
Excessive vibration
•
Excessive dust and dirt