background image

BagMixer

®

 400

6

III. CAUTION

The unit must be installed upright, on a stable, horizontal surface; never run the unit 
in any other position! Always use an earthed plug.

The housing should only be removed by qualified and duly authorized staff (see 
Chapter IX).  Never open the housing of the machine without first unplugging the unit 
from the electric mains.
In normal use, do not run the 

BagMixer

®

 400

 for more than 15 minutes continuously

(depending on sample size and nature, paddle position and speed selection). 

IF YOU ARE NOT SURE, PLEASE CONSULT US FOR ADVICE BEFORE USING!

IV. STERILE BAGS

You may use all good quality standard blender bags with 

BagMixer

®

 400

.  However, 

we strongly recommend 

interscience

 bags:

BagFilter

®

:

  bags with non-woven side filter. Highly robust.  Excellent for general use. 

BagPage

®

 bags with a full-width microperforated filtering membrane.

BagLight

®

:

  plain blender bags without filter, made in PolySilk

®

.

V. BAGMIxER

® 

400 ACCESSORIES (2)

As well as the drip tray, 

interscience

 proposes a full range of accessories to 

improve user working conditions and to increase safety and analysis quality :

BagOpen

®

:  bag holder for use when filling the bag

BagRack

®

:  storage racks for bags

BagClip

®

closing clips for blender bags

BagPipet

®

:  fixed-volume pipette (0.1 - 0.9 - 1ml)

BagTips

®

sterile straws for use with the BagPipet

®

BagSeal

®

:   heat-sealing unit for bags

Ideal complements : 

Gravimat

®

 and 

BabyGravimat

®

 

gravimetric dilutors

(2) See BagTools and BagSystem

®

 on www.interscience.fr

Summary of Contents for BAGMIXER 400

Page 1: ...BAGMIXER 400 User s manual guarantee manuel BagMixer 400 V2 2 10160 uk V2 2 UK...

Page 2: ...tifunction digital screen Variable speed 4 10 strokes sec Variable blending time 1sec 1h or open running count down Adjustable blending pressure with on screen indication Q Tight side by side stopping...

Page 3: ...an excellent preparation of solid samples before their analysis avoids the use of mixers or high speed blenders which damage the cells and bacteria contained in the sample and which require fastidious...

Page 4: ...switch at the back of the unit is in position I 2 Adjust the Timer button on top of the unit to the required setting 3 Open the front door by pulling the lever up Place a bag containing the sample in...

Page 5: ...the crank at the back of the unit chapter VII Check the settings with the GAP screen on the keyboard 5 Open the front door by pulling the lever up Place a bag containing the sample in the unit and clo...

Page 6: ...bags with BagMixer 400 However we strongly recommend interscience bags BagFilter bags with non woven side filter Highly robust Excellent for general use BagPage bags with a full width microperforated...

Page 7: ...n the door then unhook the lower edge of the door from the axis bar Swing the lower edge of the door upwards and lay the door flat on top of the unit Click the paddles and remove them 3 Fig 2 Clean th...

Page 8: ...uld only be carried out exceptionally by authorized and qualified staff Disconnect the unit from the mains then 1 Lift the lever to open the door then unhook the lower edge of the door from the axis b...

Page 9: ...BagFilter BagPage or BagLight bags from interscience Bags often leak during the cycle or at the end of the blending when blending hard wood fibers or heavy material If necessary increase the dead spa...

Page 10: ...rs if the end user returns the card within 8 days following the purchase This guarantee covering parts and workmanship will be ensured by calling your dealer or faxing interscience on 33 0 1 34 62 43...

Page 11: ...or appreciations Just let us know and we will take it into account Validate your guarantee online www interscience fr guarantee bagmixer php or by fax to 33 0 1 34 62 43 03 or by postal mail BagTools...

Page 12: ...61 Fax 33 0 1 34 62 43 03 e mail info interscience fr web www interscience fr 30 Chemin du Bois des Arpents 78860 St Nom la Bret che France BagMixer 400 GUARANTEE Thank you for choosing BagMixer 400...

Page 13: ...y interscience BagLight 400 Plain sterile bags Made of PolySilk extremely robust For samples that do not require any filtration RollBag Sampling bags For collecting sampling preservation of any sample...

Page 14: ......

Page 15: ...the end of their lifetime by our company no return carriage costs with be borne by Interscience ST NOM LA BRETECHE March 4 th 2010 Emmanuel JALENQUES Director 30 Chemin du Bois des Arpents 78860 ST NO...

Page 16: ......

Reviews: