background image

9

BagMixer

®

 400

Ix. TROUBLE-SHOOTING

x. TECHNICAL DATA

BagMixer

® 

400  is delivered with:  FREE bag samples, power cord and instruction manual.

PROBLEMS

CHECK

Unit does not start

- Power supply is correctly connected
- Main switch (at the back of the unit) is in   
   position “I” (ON)
- Door is closed correctly
- Fuse is OK
- Voltage of power supply is correct

Door closes incorrectly

Check that it is correctly positioned on the 
axis bar at the bottom of the door

Blending time is wrong

Adjust the timer

Blending speed is wrong

Adjust the speed function

Bags leak during the cycle, or at the end 
of blending

Use good quality bags: BagFilter

® 

, BagPage

®

 

or BagLight

®

 bags from 

interscience

Bags often leak during the cycle or at the 
end of the blending, when blending hard 
(wood, fibers...) or heavy material

If necessary, increase the dead space 
between door and paddles from and/or 
reduce blending speed

BagMixer

®

 400

P

W

VW

CC

®

Useful volume

50 - 400 ml

50 - 400 ml

50 - 400 ml

50 - 400 ml

Power

230/100V - 50/60Hz

230/100V - 50/60Hz

230/100V - 50/60Hz

230/100V - 50/60Hz

Size (l x w x h)

39 x 26 x 29 cm

39 x 26 x 29 cm

42 x 26 x 26.5 cm

42 x 26 x 26.5 cm

Weight

16.5 kg

16.5 kg

16.5 kg

16.5 kg

Blending time

30-210 sec. & open 

running

30-210 sec. & open 

running

1sec.-1H. and open 

running/countdown

1sec.-1H. and open 

running/countdown

Blending speed

8 strokes/sec.

8 strokes/sec.

4-10 stroke/sec.

4-10 stroke/sec.

Drip tray

option

Reference

021 230

022 230

023 230

024 230

Summary of Contents for BAGMIXER 400

Page 1: ...BAGMIXER 400 User s manual guarantee manuel BagMixer 400 V2 2 10160 uk V2 2 UK...

Page 2: ...tifunction digital screen Variable speed 4 10 strokes sec Variable blending time 1sec 1h or open running count down Adjustable blending pressure with on screen indication Q Tight side by side stopping...

Page 3: ...an excellent preparation of solid samples before their analysis avoids the use of mixers or high speed blenders which damage the cells and bacteria contained in the sample and which require fastidious...

Page 4: ...switch at the back of the unit is in position I 2 Adjust the Timer button on top of the unit to the required setting 3 Open the front door by pulling the lever up Place a bag containing the sample in...

Page 5: ...the crank at the back of the unit chapter VII Check the settings with the GAP screen on the keyboard 5 Open the front door by pulling the lever up Place a bag containing the sample in the unit and clo...

Page 6: ...bags with BagMixer 400 However we strongly recommend interscience bags BagFilter bags with non woven side filter Highly robust Excellent for general use BagPage bags with a full width microperforated...

Page 7: ...n the door then unhook the lower edge of the door from the axis bar Swing the lower edge of the door upwards and lay the door flat on top of the unit Click the paddles and remove them 3 Fig 2 Clean th...

Page 8: ...uld only be carried out exceptionally by authorized and qualified staff Disconnect the unit from the mains then 1 Lift the lever to open the door then unhook the lower edge of the door from the axis b...

Page 9: ...BagFilter BagPage or BagLight bags from interscience Bags often leak during the cycle or at the end of the blending when blending hard wood fibers or heavy material If necessary increase the dead spa...

Page 10: ...rs if the end user returns the card within 8 days following the purchase This guarantee covering parts and workmanship will be ensured by calling your dealer or faxing interscience on 33 0 1 34 62 43...

Page 11: ...or appreciations Just let us know and we will take it into account Validate your guarantee online www interscience fr guarantee bagmixer php or by fax to 33 0 1 34 62 43 03 or by postal mail BagTools...

Page 12: ...61 Fax 33 0 1 34 62 43 03 e mail info interscience fr web www interscience fr 30 Chemin du Bois des Arpents 78860 St Nom la Bret che France BagMixer 400 GUARANTEE Thank you for choosing BagMixer 400...

Page 13: ...y interscience BagLight 400 Plain sterile bags Made of PolySilk extremely robust For samples that do not require any filtration RollBag Sampling bags For collecting sampling preservation of any sample...

Page 14: ......

Page 15: ...the end of their lifetime by our company no return carriage costs with be borne by Interscience ST NOM LA BRETECHE March 4 th 2010 Emmanuel JALENQUES Director 30 Chemin du Bois des Arpents 78860 ST NO...

Page 16: ......

Reviews: