D-0120588-B
–
2019/03
Eclipse
–
Instructions for Use - EN
Page 91
Temporary setup
allows for making temporary changes to the
selected protocol. The changes will be valid for the current session
only. After making the changes and returning to the main screen,
the name of the protocol will be followed by an asterisk (*).
List of historical sessions
accesses historical sessions for review
or the
Current Session
.
The
historical session
box can be expanded by dragging
downwards with the mouse or minimized/maximized by clicking on
the arrow button.
The session highlighted in orange, is the selected session
displayed on screen. Check the checkbox next to the session date
to
overlay historic sessions
on the graph.
Go to current session
takes you back to the current session.
Probe status
is shown by a colored bar with description next to it.
When the probe status is
Out of ear
it will show the color of the
selected ear (blue for left and red for right). When the probe is
detected to be
In ear
the color is green. When
Blocked
,
Leaking
or
Too Noisy
the colored bar is amber. When
No probe
is
detected the status bar is grey.
Summary view
toggles between displaying result graph or result
graph with test summary tables.
Monaural/Binaural view
toggles between showing the results of a
single ear or both ears.
Report editor
button opens a separate window for adding and
saving notes to the current or historic session.
Checking the
Acceptable noise level Off
box disables the
rejection of any recording even when there is too much noise in the
recording.
The
Acceptable noise level
slider allows setting the acceptable
noise level limit b30 and +60 dB SPL. Sweeps recorded
above the set acceptable noise level are considered to be too noisy
and are rejected.
The VU meter indicates the current noise level and becomes amber
in color when exceeding the set level.
The hardware indication picture
indicates whether the
Lyra or
Eclipse hardware is connected.
Simulation mode
is indicated when operating the software in
simulation mode (user chosen).
Before testing, the
Timer
symbol indicates after how much time the
TEOAE test will automatically stop. During testing the timer counts
down to zero. You can disable the countdown by clicking on the
timer during testing. As a result the timer will begin counting up and
indicate how much test time has passed. Then the test will continue
until you manually press stop.
When the measurement is rejecting, the timer will stop counting.
Artifact rejection
is dependent upon the
Acceptable Noise Level
setting and
Level tolerance
level set in the protocol.
Summary of Contents for Eclipse ABRIS