D-0120588-B
–
2019/03
Eclipse
–
Instructions for Use - EN
Page 125
11
Technical specifications
11.1 Technical specifications - Eclipse hardware
Medical CE-mark:
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II of the
Medical Device Directive 93/42/EEC.
Approval of the quality system is made by TÜV
–
identification no. 0123.
Standards:
Safety:
IEC 60601-1:2012, Class I, Type BF
CSA C22.2 No.60601-1:2014 and US ES 60601-1:2012.
EMC:
IEC 60601-1-2:2014
Power Supply:
Input Volts:
100
–
240VAC, 50/60Hz.
Consumption:
26W (0.3A Max)
Safety marking
ECM60UT31-XE0410 and cURus
Operating environment:
Operating Temperature:
15
–
35
C (59 - 95
F)
Rel. Humidity:
30
–
90%
Ambient Pressure:
98kPa
–
104kPa
Transport & Storage:
Storage Temperature:
Transport Temperature:
Rel. Humidity:
0
C
–
50
C (32°F - 50°F)
-20
–
50
C (-4°F - 122°F)
10
–
95% (non condensing
Warm up time:
10 minutes at room temperature (20
C) (68
F).
General
PC control:
USB:
USB 1.1 or 2.0 for input/output for computer communication.
Eclipse if fully operated from a PC
Construction:
Metal cabinet
Edipse Dimensions
(L x W x H) 28 x 32 x 5.5 cm (11 x 12.6 x 2.2 Inches)
Eclipse Weight
2.5kg / 5.5 lbs excluding accessories
Summary of Contents for Eclipse ABRIS
Page 1: ...Instructions for Use EN Eclipse Science made smarter D 0120588 B 2019 03...
Page 2: ......
Page 80: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 72...
Page 112: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 104...
Page 132: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 124...