9.8.6 Extending the test time
Right click on the
Right
or
Left
text above the Test Frequency Graphs to select Extend Test Time for all test
frequencies for the selected ear.
Right click within the desired Test Frequency Graph to Extend Test Time for the selected frequency.
9.8.7 Adjusting the stimulus intensity
Right click on the
Right
and
Left
text above the Test Frequency Graphs to select and initiate an overall
stimulation intensity for the entire selected ear immediately. Ongoing recordings for the selected ear will be
stopped.
Right click within the desired Test Frequency Graph to change the Stimulus Intensity for the selected
frequency. Only intensities within the grey area can be selected. This safeguards against stimulating at
intensity levels greater than a 20dB difference between the test frequencies.