4.6 Making a cVEMP recording
The cVEMP test is a test to assist in the assessment of vestibular function.
4.6.1 Electrode montage for cVEMP
Red
Right SCM (reference or inverting)
White
Left SCM
Black
Ground
Blue
Clavicular joint (active or non-inverting)
4.6.2 Stimuli for cVEMP
Typically, a 500Hz tone burst is used for cVEMP testing as it provides the largest response amplitude. Other
stimuli available for testing include:
•
Tone bursts 250Hz
–
4kHz
•
Click
•
Custom WAV file (when Research Module license is enabled).