TENSPros
Guidance and- manufacturer’s declaration. Electromagnetic immunity
The device is intended for use in. the electromagnetic environment specified
below. The customer or the user should assure that it is used in such an
environment.
Immunity
test
IEC 60501
test level
Complia
nce
level
Electromagnetic environment -
guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted
RF lEC
61000-4-6
3 Vrms 150
kHz to 80
MHz
3 Vrms
Radiated
RF lEC
61000-4-3
3 V/m80 MHz
to 2.5 GHz
3 V/m
, 80MHz to 800MHz
, 800MHz to 2,5MHz
Where P is the maximum output power
rating of the transmitter In watts (W)
according to the. Transmitter
manufacturer and d Is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur In the vicinity
of equipment marked with the following
symbol:
NOTE I At 80 MHz ends 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
Summary of Contents for IF COMBO DI4738
Page 1: ...T E N S P r o s if combo INSTRUCTION MANUAL ...
Page 8: ...T E N S P r o s ...
Page 11: ...T E N S P r o s To set the button and hold for To set the 2 2 LCD display ...
Page 33: ...T E N S P r o s ...
Page 34: ...T E N S P r o s ...
Page 35: ...T E N S P r o s ...