10
EXTERNAL CALIBRATION *
* (For all models except Analytical
balance).
The electronic reading balance takes
measurement of mass making use of
gravity (g). Difference in geographic area
and altitude will vary gravity acceleration
(g). Therefore, for accurate measurement,
scale must be adjusted to the local
environment. This adjustment is
accomplished by
CAL
key.
At empty pan, press
CAL
key.
Dashes are displayed.
c a l i b . w e i g h t
B
EL
ENGINEERING
2 0 0 . 0 0
lb
oz
ct
g
The display will stop flashing, indicating
calibration weight value.
Wait for buzzer BEEP to confirm
calibration.
Unload calibration weight from the pan.
The balance is ready for weighing
operations.
NOTE:
if there is an interference during
calibration process, the message
“ERR
OR”
B
EL
ENGINEERING
- - - - - - - -
g
lb
ct
oz
B
EL
ENGINEERING
2 0 0 G r
lb
When calibration weight value starts to
flash, load the weight on the pan.
B
EL
ENGINEERING
0 . 0 0
lb
ct
oz
g
CAL
c a l i b .
w e i g h t
Intell-Lab
™
Intell-Lab
™
Intell-Lab
™
Intell-Lab
™
Summary of Contents for intell-lab ph series
Page 4: ...WEIGHING PAN ASSEMBLY June 2009...
Page 5: ...WEIGHING PAN ASSEMBLY 5...
Page 11: ...11 will be displayed...
Page 15: ...15 starting from point 2...
Page 19: ...19 ons or draughts...
Page 28: ...28 Instructions 6K UK Rev 08 doc Page 23 of 46...
Page 39: ...39 Possible non meaningful zeroes are spaces...
Page 41: ...41 alphanumeric keyboard...
Page 46: ......