)Moreover, it is possible to calibrate the
balance with a calibration weight higher
than the one set.
To perform this sort of calibration,press
and keep CAL key pressed at empty pan
until the acoustic alarm will be
deactivated, then release the key and the
following string " -------- " will be displayed,
followed by the flashing message "LOAD"
.
Now load on the pan a weight equal or
higher than calibration weight set; the
balance will recognise as valid a weight
equal or higher than calibration weight as
long as it is a whole number in
comparison with the most meaningful
digit of calibration weight..
Example: if calibration weight is 200g, it
will be possible to calibrate the balance
with values from 200g,300g,400g up to
the highest limit of balance weighing
range.
The message
"LOAD"
on display will
stop flashing.
Once calibration has been effected, the
value of calibrated weight will be
displayed.
0 . 0 0
Remove calibration weight.