33
AUTOMATIC UPDATING OF
UNIT WEIGHT
After having effected the sampling, it is
possible to update the average unit weight
as follows :
instead of loading all the pieces to count,
load a number of pieces equal to the
double of the one loaded on the pan.
Wait for all the arrows to be displayed
followed by a short acoustic message.
Now it is possible to repeat this procedure
up to a maximum of 255 pieces or proceed
to the normal counting of the pieces. This
mechanism grants a more precise
evaluation of the average unit weight and a
better precision in counting of pieces.
NOTE:
the mechanism of the automatic
updating is not active if the sampling has
been effected by insertion of the average
unit weight from alphanumeric keyboard.