M L X 3 0 0 w a t t X e n o n L i g h t s o u r c e
OPERATION
XENON LAMP MODULE REPLACEMENT
5. the system will start up at the same light intensity as when last used (Figure 3). For new systems, the light
intensity will start at the minimum 20% setting.
6.to adjust the light intensity, push the Membrane switches (Figure 3): + to increase or - to decrease the light
output or intensity. range is 0%, 20 – 100% in 5% increments. holding the + or – button will change the intensity
more quickly.
note: it is strongly recommended that the light be used at the minimum intensity for good visualization.
7. the system status Display can be pressed to show both lamp time and system operation time. a second press
of the switch will show the software version of the system monitor. a third press returns to light output. (the
system will automatically return to the light intensity reading after15 seconds).
8. the system can be placed in stand-by mode (no light emitted) allowing the surgeon to unplug a headlight and
move around the table, or to change cables in a laparoscopic or endoscopic procedure.
CAUTION: Cable end fitting can be HOT!!!
note: if unit is left in stand-by mode for 15 minutes, it will automatically shut off the lamp. Fans will keep running.
to re-activate, push the stand-by Button.
9. to shut the system off, simply press the main Power switch as in step 2.
Note: Please adhere to appropriate safety precautions when performing lamp replacement. Only qualified
personnel should service this device. Protective facemask and/or proper safety glasses should be
worn when replacing the lamp module. Before changing the lamp module, turn power off and allow
the light source to cool for at least fifteen (15) minutes.
Please read and comply with all Precautions and general warnings listed in this manual.
Figure 3
system
status
Display
Membrane
switches