9.5. VENTILATION PARAMETERS
The Ventilation Parameters screen displays the relevant patient information,
updated on a real-time basis:
Ventilation Parameters screen
•
Select
BPM
to change the breaths per minute delivered.
•
Select
VTe
to change the desired VTi.
•
Select
I:E
to change the default T.i
•
Select
ALARMS
to view the current active (silenced) alarms.
•
Select
GRAPH
to display the Pressure or Flow graphs – see Section
•
Select
MANUAL BREATH
to give one breath to the patient using the
current parameters.
•
Select
MENU
to display the Menu screen – see Section
Note:
The possible range of T.i is a minimum of 0.3 seconds and max-
imum of 4 seconds. However, a maximum will be restricted in accor-
dance with the breath length (as defined by the BPM) to a maximum of
half the breath time. For example, if the BPM is defined as 20, then the
breath length will be 3 seconds and the max T.i will be 1.5 seconds).