Ventway Sparrow
User Manual
Page 78 of 144
12. DEFAULT PARAMETERS
12.1. START VOLUME VENTILATION
Patient Weight Rate Tidal Volume
(Kg)
(bpm)
(ml)
5
35
50
10
30
70
15
23
100
20
20
120
30
18
180
40
16
240
50
14
300
60
12
360
70+
12
450
12.1.1. OXYGEN SUPPLY
When a high-pressure oxygen source that is connected to an Oxygen Mixer is
not available, the Ventway Sparrow ventilator can accept oxygen from a low-
pressure oxygen source such as a reservoir bag connected to a flow meter.
To do this, use an optional low-pressure oxygen enrichment system attached
to the ventilator air inlet port through an optional Ventway adapter.
Adjust the “O
2
coefficient” parameter on the device through the O
2
enrichment screen, which can be found under the
VENT. PARAMS
screen, so
that the coefficient is aligned with the actual FiO
2
value given to the patient.
Select the "O
2
ENRICHMENT" option in the
VENT. PARAMS
menu.
The FiO
2
value must be measured with a calibrated external oxygen analyzer.
If a high-pressure oxygen source is available, an Inovytec-approved Oxygen
Mixer can be used to connect to the air inlet port, delivering between 30% to
95% FiO
2
to the patient.
Measure FiO
2
with a calibrated external oxygen
analyzer.
The oxygen supply pressure shall be according to manufacturer specifications
(usually 40-60 psi).
12.1.3. Additional recommended legally marketed components /
for recommended legally marketed reservoirs.
Summary of Contents for Ventway Sparrow
Page 27: ...Ventway Sparrow User Manual Page 27 of 144 5 5 PNEUMATIC SECTION THEORY OF OPERATION...
Page 84: ...Ventway Sparrow User Manual Page 84 of 144 VWSP 100MR Civil Model VWSP 900 MR Robust Model...
Page 91: ...Ventway Sparrow User Manual Page 91 of 144...
Page 94: ...Ventway Sparrow User Manual Page 94 of 144...
Page 95: ...Ventway Sparrow User Manual Page 95 of 144...