Ventway Sparrow
User Manual
Page 114 of 144
19. BATTERIES
The Sparrow can be configured with two battery options: an internal, re-
chargeable battery pack, and an internal non-rechargeable battery pack.
The battery should be charged or replaced when not in use. For long
storage periods, the battery should not be inserted into the battery
compartment of the ventilator. It is possible to switch between these two
battery options.
19.1. BATTERY MAINTENANCE
Rechargeable batteries
The internal battery pack is charged from the external device connector,
using the supplied 18-24 volt regulated power supply.
Caution:
If the batteries do not charge, replace the charger or the
batteries.
Caution:
Battery failure may occur if maintenance is not performed on
time and according to instructions.
Note:
If charging at temperatures over 35°C (95°F), charging time may be
extended, and battery thermal protection may be activated preventing
the battery charge.
Shelf life: 4 years. First time charging: 6 hours.
Permissible battery pack charging temperature range: 0°C to 50°C (32°F
to 122°F)
Permissible temperature range for long-term storage: 10°C to 35°C (50°F
to 95°F)
Charge Interval: 6 months in standard temperature and pressure
Shelf life: 4 years or 500 cycles at 25°C (77°F). A cycle is defined as one
charge and one discharge.
Max discharge current: 8.25[A]
Storage temperature for shipping state (about 40% capacity of fully
charged state):
1 month
45°C to 60°C (113°F to 140°F)
3 month
-25°C to 45°C (-13°F to 113°F)
1 year
-20°C to 25°C (-4°F to 77°F)
Non-Rechargeable batteries
Shelf life: 5 years. Long term storage: -20°C to 35°C (-4°F to 95°F)
Summary of Contents for Ventway Sparrow
Page 27: ...Ventway Sparrow User Manual Page 27 of 144 5 5 PNEUMATIC SECTION THEORY OF OPERATION...
Page 84: ...Ventway Sparrow User Manual Page 84 of 144 VWSP 100MR Civil Model VWSP 900 MR Robust Model...
Page 91: ...Ventway Sparrow User Manual Page 91 of 144...
Page 94: ...Ventway Sparrow User Manual Page 94 of 144...
Page 95: ...Ventway Sparrow User Manual Page 95 of 144...