18.5. VENTILATION PERFORMANCE
Note:
Measurement accuracy of calibration equipment may affect
the respiratory parameters display by the following parameters:
•
Volume calculation: +/-2% or 20 [cc]
•
Inspiratory and expiratory times: +/- 0.02 [s]
•
Pressure: +/- 0.75% or +/- 0.1 [mbar]
Note:
If temperature exceeds 85°C, a temperature alarm may be
activated. This may result in breath parameters inaccuracies that
exceed the tolerances specified below.
Note:
If humidity exceeds 95% and condensation occurs, it may lead
to water droplet buildup in the sensing tubes.
Note:
If the ventilator is exposed to altitudes outside of the defined
values, the ventilator may be unable to accurately compensate for
volume and pressure deviations.
Respiratory Rate
1 – 60 ± 1 bpm
Tidal Volume
50-2000 mL
Accuracy of
Respiratory Volume
Measurement
±10% + 4 ml
Inspiratory Pressure
Limit
11 to 60 cm H
2
O
Inspiratory Time
0.3 to 4 ±10% sec
I:E Ratio
1:4 to 4:1
Peak Flow (PIF)
135 (-5% to +10%)% L/min
Spontaneous to 160 ±10% L/min
Oxygen Mix (FiO
2
)
21% to 95% ±5% O
2
PEEP
0 to 20 cm H
2
O ±1, (externally adjusted)
Trigger Sensitivity
-0.25 to -10 cm H
2
O Pressure, Off
PS
5 to 50 cm H
2
O
Controlled Pressure
5 to 60 cm H
2
O
FiO
2
low pressure
enrichment
21% to 100% (low and high pressure oxygen
source)
Summary of Contents for Ventway Sparrow
Page 27: ...Ventway Sparrow User Manual Page 27 of 144 5 5 PNEUMATIC SECTION THEORY OF OPERATION...
Page 84: ...Ventway Sparrow User Manual Page 84 of 144 VWSP 100MR Civil Model VWSP 900 MR Robust Model...
Page 91: ...Ventway Sparrow User Manual Page 91 of 144...
Page 94: ...Ventway Sparrow User Manual Page 94 of 144...
Page 95: ...Ventway Sparrow User Manual Page 95 of 144...